Mountz Torque LTT Torque Analyzer Operation Manual

The display holds the first detected torque peak applied. Before any torque is applied, the display show dashes in the torque value area. Once peak is detected, the display will show the torque value. If a second peak is detected then it will be displayed in the lower right (in small reverse video). First Peak

Selecting First Peak

1. Press the "Menu" Button. 2. Use the Scroll Down key and select Setup by highlighting and pressing Enter . 3. Use the Scroll Up or Down key to toggle through the Setup options. 4. Press Enter key to select a setup setting for First Peak .

Tolerance Auto Clear Angle Setup T+A Threshold First Peak Spindle Audit

Filter Units Mode Setup

1. Use the Scroll Up or Down key to toggle through the sensitivity settings: (Low, Medium &High) 2. Press the Right or Left key to move down to the Min Peak setting location 3. Use the Scroll Up or Down to set the Minimum Peak 4. Press Enter key once setting is complete. 5. Press Main button to return to main display screen.

First Peak Setup

Sensitivity: Low Min Peak: 5.0089 Lbf.in

Main

Page 13