External Connections USB
The computer connection is USB. There is no setup required. This allows for data to downloaded to a PC. The PC will require a USB I/O.
USB
External Transducer Input
Power Supply Port
External Transducer Input The transducer connector is a high density D-Sub connector with 15 pins. The pin description is shown below:
Pin Number
Function
Description
1 2 3 4 5 6 7 8 9
Analog Ground Brushless Signal Multiplexer Sensor Direct Digital Ground Excitation Voltage +
Output from brushless torque sensors + /- 5V
Used for the Multiplexer Used to detect torque sensors
5V or 16 volts depending on bridge or brushless torque sensor
Bridge Signal - Multiplexer Angle Lead Sensor Drive Bridge Signal +
Negative output from bridge torque sensor
Used for the Multiplexer
TTLoutput from angle detector in angle sensor
10 11 12 13 14 15
Used to detect torque sensors
Positive output from bridge torque sensor
Shield
Angle Trail
TTL output for angle lags Angle Lead by 90 degrees
+5 Volt Angle Supply
Data for “Smart” torque sensors Proprietary ARCII protocol for smart torque sensors from Mountz
User Interface
Display Screen Displays the menu structures, torque readings, operating mode, torque units.
Function Keys Functions will be identified on the display screen.
Go and No Go LEDs Used to monitor lower and upper torque limits and receive a visual warning.
Internal Transducer Built-in transducer
Enter
Scroll Up
“Scroll and Enter Keys” Used to toggle through the different menu structures.
Scroll Left
Scroll Right
Scroll Down
Page 3
Corporate Headquarters: 1080 North 11th Street, San Jose, CA95112 Phone: (408) 292-2214 Fax: (408) 292-2733 www.mountztorque.com
Made with FlippingBook Annual report maker