Mountz Torque LTT Torque Analyzer Operation Manual

External Connections USB

The computer connection is USB. There is no setup required. This allows for data to downloaded to a PC. The PC will require a USB I/O.

USB

External Transducer Input

Power Supply Port

External Transducer Input The transducer connector is a high density D-Sub connector with 15 pins. The pin description is shown below:

Pin Number

Function

Description

1 2 3 4 5 6 7 8 9

Analog Ground Brushless Signal Multiplexer Sensor Direct Digital Ground Excitation Voltage +

Output from brushless torque sensors + /- 5V

Used for the Multiplexer Used to detect torque sensors

5V or 16 volts depending on bridge or brushless torque sensor

Bridge Signal - Multiplexer Angle Lead Sensor Drive Bridge Signal +

Negative output from bridge torque sensor

Used for the Multiplexer

TTLoutput from angle detector in angle sensor

10 11 12 13 14 15

Used to detect torque sensors

Positive output from bridge torque sensor

Shield

Angle Trail

TTL output for angle lags Angle Lead by 90 degrees

+5 Volt Angle Supply

Data for “Smart” torque sensors Proprietary ARCII protocol for smart torque sensors from Mountz

User Interface

Display Screen Displays the menu structures, torque readings, operating mode, torque units.

Function Keys Functions will be identified on the display screen.

Go and No Go LEDs Used to monitor lower and upper torque limits and receive a visual warning.

Internal Transducer Built-in transducer

Enter

Scroll Up

“Scroll and Enter Keys” Used to toggle through the different menu structures.

Scroll Left

Scroll Right

Scroll Down