Mountz Torque LTT Torque Analyzer Operation Manual

Note for Force & Load Sensors When performing a Load Cell Calibration, if choosing a 2 point calibration, load at 10% and 100%. If performing a 3 point calibration and load at load at 10%, 50% and 100%. The calibration program will guide you through the calibration by text in the information box. By selection in the Calibration Type you can select Compression Only, Tension Only or Both if the Load cell supports this. Type in the appropriate Full Scale value and select the units desired. The "Delete Device" option allows the ability to delete non-smart sensors or transducers records from the LTTTorque Analyzer. In order to delete the device record for a non-smart sensor, the non smart sensor or transducer must be con- nected to the LTT Torque Analyzer. Ensure the LTT Torque Analyzer is powered on. Connect the non-smart device to the LTT. Use the scroll up or scroll down arrows on the keypad of LTTto locate the device to be deleted. Press enter on the LTTkeypad to complete the selection. Launch the torque meter interface program and go to the Sensor Calibration tab. Press "the Read From Sensor" tab located just above the "Delete Device" tab to read in the device information. A Calibration Data window should appear with details of the non-smart device. Click on exit to exit the Calibration Data win- dow. Please ensure to validate the details displayed on the screen to ensure the selection is as desired, as the "Delete Device" function will delete the device record. A non-smart device can only be added back in by going through the cali- bration process. Press the "Delete Device" tab to remove the device record for the non-smart device. If there is a need to delete multiple non-smart devices, the above steps must be repeated for each non-smart device.

Page 31

Corporate Headquarters: 1080 North 11th Street, San Jose, CA95112 Phone: (408) 292-2214 Fax: (408) 292-2733 www.mountztorque.com

Made with FlippingBook Annual report maker