Test

Fluke 154 HART Calibration Assistant

Fluke 154 Application

HART Calibration assistance at your fingertips. Combine with a Fluke Calibrator for a total HART Callibration Solution. The 154 puts HART device configuration at your fingertips. An Android™ based tablet user interface makes HART configuration easy. The external wireless modem allows you to connect the modem to the transmitter being tested so you can communicate to the device remotely. HART transmitters can often be found in confined spaces that are difficult to access, or are out of reach. The Fluke 154 helps reduce the need to stand right next to the device being tested or configured so you can work from a safer, more convenient location. When you combine the Fluke 154 Communicator with a Fluke 750 Series Documenting Process Calibrator, or a Fluke 720 Series Multifunction Process Calibrator you have a complete HART calibration and configuration solution. • Full HART Device Description (DD) support of all HART devices • Configure HART devices • Perform HART trim on HART devices when used with a Fluke 750 or 720 Series Calibrator • Monitor device PV, SV, TV, QV and other measured HART variables • Quarterly HART DD updates for free • Convenient wireless connectivity to HART modem • Easy to use, fast connect and view HART data • Store HART device configurations • Long range wireless communication up to 76.2 metres • Three year warranty

Fluke 154

Tablet Operating System

Android 5.0 or newer

Processor

ARM or Atom

Memory RAM Memory ROM

1 GB 2 GB

Screen

960x540 qHD

Communication Port HART Modem Enclosure Dimensions (LxWxH) Total Unit Weight HART Cable Length HART Cable Connectors

Bluetooth

10.16 cm x 2.5 cm x 6.9 cm

0.16 kg

1.2 m

Mini-Grabber

HART Interface Interface

Capacitive coupling

Output

600 mVpp

Leakage

< 10 uA

Connections Specifications

Across loop resistor or HART field device HART 4, HART 5, HART 6, HART 7

HART Physical Layer Spec

HCF_SPEC-54 Bluetooth v2.0

Specification

Included Accessories Tablet Charger and USB, Micro-cable, Fluke HART App (installed), HART Modem, USB Mini-Cable, Hook Test Clips, Alligator Clips, Magnetic Hanging Strap, Case, Manual installed on Tablet.

Frequency

2.4 GHz

Class 1 Master Distance

275 ft (83.8 m)

Battery Type Battery Life

Rechargeable Li-ion

19 hours continuous use

Humidity

0 % to 90 % (non-condensing)

Ordering Information

HART Calibration Assistant

Fluke 154

C ontact your local distributor for price and availability. www.fluke.com/au/wtb

110

Process Calibration Tools

Made with FlippingBook Learn more on our blog