Fluke Calibration 9132/9133 Portable Infrared Calibrator
Recommended Accessories
Precision when you need it for infrared temperature calibration Whether you’re using in-line or handheld infrared pyrometers, you need good calibration standards to verify their accuracy. Our portable IR calibrators provide stable blackbody targets for calibrating noncontact IR thermometers from –30 °C to 500 °C. These units feature a temperature controlled measurement surface with a diameter of 57 mm. The target temperature can be controlled in set-point increments of 0.1 ° from –30 °C to 500 °C and a well is located directly behind the blackbody surface for contact calibration. • Certify IR pyrometers from –30 °C to 500 °C • Large 57 mm blackbody target • RTD reference well for contact temperature measurement • Small, compact design • One year warranty
9308 Carrying Case
9302 Carrying Case
Fluke 9132
4180-APRT 2” Aperture, (4180 or 4181)
4180-CASE Carrying Case, (4180 or 4181)
Fluke 9133
General Specifications
9132
9133
Temperature Range
50°C to 500°C ±0.5°C at 100°C ±0.8°C at 500°C
–30°C to 150°C at 23°C ambient
Accuracy
±0.4°C
Target Size
57 mm
57 mm
Target Emissivity
0.95 (±0.02 from 8 to 14 mm) 30 minutes (50°C to 500°C) 30 minutes (500°C to 100°C)
0.95 (±0.02 from 8 to 14 mm) 15 minutes (25°C to 150°C) 15 minutes (25°C to –20°C)
Heating Time Cooling Time
Fluke Calibration 4180/4181 Precision Infrared Calibrators
Accredited performance for point and shoot calibrations Now it’s easy to increase your IR temperature measurement accuracy with the 4180 and 4181 Precision Infrared Calibrators. These infrared temperature calibrators give more consistent, accurate and reliable calibrations because emissivity is radiometrically calibrated, and the target size minimises size of source effect errors. In addition, these infrared calibrators simplify calibration because they uniquely compensate for errors caused by thermometer emissivity settings. • Calibrated radiometrically for meaningful, consistent results • Accurate, reliable performance from –15 °C to 500 °C • Large target size of 152 mm is required for calibrating most thermometers • Accredited radiometric calibration report • One year warranty
Ordering Information
Fluke 4180
9312-256
Portable IR Calibrator 50°C to 500°C Portable IR Calibrator -30°C to 150°C
9133-256
4180
Precision Infrared Calibrator, –15 °C to 120 °C Precision Infrared Calibrator, 35 °C to 500 °C
4181
For more information, go to http://au.flukecal.com
www.fluke.com/au/PCT
Process Calibration Tools
115
Made with FlippingBook Learn more on our blog