Catalog 91 - Force And Material Testing

L oad C ell S ensors

Starrett offers a full range of precision load cell sensors for material testing, force analysis and force measurement applications. Starrett load cells are compliant with IEEE 1451.4 and meet or exceed ASTM E4, BS 1610, ISO 7500-1 and EN 10002-2. Measurement accuracies of ±0.05% of reading down to 1/100 of sensor capacity may be achieved. Sensors are supplied with a NIST-traceable Certificate of Calibration. BLC L oad C ell S ensors Starrett BLC load cell sensors are full-bridge, temperature compensated, strain gage instruments designed and optimized for basic force testing applications. These S-beam sensors feature high axial stiffness and minimal deflection at full capacity which leads to improved measurement accuracy. The BLC sensors are general purpose sensors available in capacities from 2lbf to 500lbf (10 to 2500N). These sensors are used exclusively with FMM motorized test frames.

BLC - Basic Force Measurement S-beam Sensors

Load Capacity

Safe Overload Full Scale Deflection Height

Width

Thread

Model No.

N

KGF

LBF

% Full Scale in

mm in

mm in

mm mm

BLC-2 BLC-5

10 20 50

1 2 5

2 5

150 150 150 150 150 150 150 150

0.009 0.008 0.007 0.007 0.006 0.003 0.003 0.005

0.22 0.21 0.18 0.18 0.15 0.08 0.08 0.13

3.0 3.0 3.0 2.0 2.0 2.0 2.0 2.0

76.2 76.2 76.2 50.8 50.8 50.8 50.8 50.8

3.0 3.0 3.0 2.0 2.0 2.0 2.0 2.0

76.2 76.2 76.2 50.8 50.8 50.8 50.8 50.8

M6 x 1-6H M6 x 1-6H M6 x 1-6H M6 x 1-6H M6 x 1-6H M6 x 1-6H M6 x 1-6H

BLC-10 BLC-20 BLC-50 BLC-100 BLC-200 BLC-500

10 20 50

100 250 500

10 25 50

110 225 550

1000 2500

100 250

M12 x 1.75-5H

NOTES Load measurement accuracy is ±0.1% of load cell capacity. Display resolution is 10,000:1. Starrett recommends on-site verification of accuracy during installation. Sensor calibration should be performed at least annually.

MLC L oad C ell S ensors The Starrett MLC load cell sensors are full-bridge, temperature compensated,strain gage instruments designed and optimized for material testing applications. These low profile sensors feature high axial stiffness and minimal deflection at full capacity which leads to improved measurement accuracy. The MLC are general purpose sensors available in capacities from 125N to 50kN. For FMS, MMS, FMD, or MMD test frames.

MLC Series - Low Profile Sensors

Load Capacity

Safe Overload Full Scale Deflection Height 1

Width

Thread

Model Number

N

KGF 12.5

LBF

% Full Scale mm inch

mm inch

mm inch

mm

MLC-125 MLC-250 MLC-500 MLC-1000 MLC-1500 MLC-2500 MLC-5K MLC-10K MLC-25K MLC-50K

125 250 500

28 56

150 150 150 150 150 150 150 150 150

0.08 0.08 0.08 0.08 0.03 0.03 0.03 0.03 0.05 0.05

0.003 0.003 0.003 0.003 0.001 0.001 0.001 0.001 0.002 0.002

38.1 38.1 38.1 38.1

1.5 1.5 1.5 1.5

69.8 69.8 69.8 69.8

2.75 2.75 2.75 2.75 4.13 4.13 4.13 4.13 4.13 4.13

M6 x 1-6H M6 x 1-6H M6 x 1-6H M6 x 1-6H M16 x 2-4H M16 x 2-4H M16 x 2-4H M16 x 2-4H M16 x 2-4H M16 x 2-4H

25 50

112 225 337 562

1000 1500 2500 5000

100 150 250 500

63.51 63.51 63.51 63.51 63.51 63.51

2.51 2.51 2.51 2.51 2.51 2.51

104.8 104.8 104.8 104.8 104.8 104.8

1124 2248 5620

10,000 1000 25,000 2500 50,000 5000

11,250 150

NOTES 1 Dimension includes the base adapter. These MLC sensors are supplied with the base adapter standard. Base adapters are recommended for any MLC sensor.

Load measurement accuracy is ±0.05% of reading down to 1/100 of load cell capacity. Display resolution is 10,000:1. Starrett recommends on-site verification of accuracy during installation. Sensor calibration should be performed at least annually.

54

Made with FlippingBook - professional solution for displaying marketing and sales documents online