Step 3:
The screen will briefly display “CAL” button to cycle through the options to reach yo select.
The screen will briefly display “CAL” before showing the torque unit selection screen. Press Lever to cycle through the options to reach your desired units, then press Calibration button it selection screen. Press Lever ur desired units, then press Calibration button again to
Torque units
Calibration button
FGA-8, FGA-20 & FGA-40 Five options available:
FGA-80 MODEL Three options available:
• ozf.in • lbf.in • cN.m • N.m • kgf.cm
• lbf.in • N.m • kgf.cm
Step 4:
Using the Lever button, cycle through the ranges to your tool model. Then press the display can be found below. Using the Lever button, cycle through the ranges until the display reads the value corresponding r tool model. Then press the Calibration button to accept. A breakdown of the models and their until the display reads the value corresponding button to accept. A breakdown of the models and their
Example Example: FGA-40
FGA-8 model: FGA-20 model: FGA-40 model: FGA-80 model:
8.00 lbf.in 20.0 lbf.in 40.0 lbf.in 80.0 lbf.in
The value displayed when selecting the m units were selected before. If you are calibrating an FGA capacity for that model in cN.m torque unit. The value displayed when selecting the model is the max range of the tool and displays on whatever units were selected before. If you are calibrating an FGA-8 in cN.m, you need to choose the maximu m torque unit. odel is the max range of the tool and displays on whatever 8 in cN.m, you need to choose the maximum
Step 5: Before starting the calibration process, used for it. The screen will bring up the first point of calibration displayed in the percentage of max setting. The image on the left reads 10%. This first percentage will differ depending on which tool model is being calibrated: Before starting the calibration process, recommend that you calculate the torque values that will be The screen will bring up the first point of calibration displayed in the percentage of max setting. The image on the left reads 10%. Calculate target torque based on the displayed p recommend that you calculate the torque values that will be torque based on the displayed percentage.
Example Example: FGA-40
FGA-8 model: 20%
FGA-20, FGA-40 and FGA
40 and FGA-80 models: 10%