prima

The PRIMA Humeral System is intended to be used with all glenoid implants belonging to the SMR Shoulder System. Instructions for use of SMR Shoulder System should be consulted for warnings that apply to these systems. The SMR Shoulder System components which are compatible with PRIMA Humeral System components in anatomical and in reverse configuration are:

System

Components

A

R

ü

SMR Humeral Heads (Standard, CTA)

ü ü

SMR TT Metal Back (Baseplate and Peg)

ü

SMR Axioma Metal Back Liner

ü ü

SMR Axioma TT Glenoid (Baseplate and Peg)

ü ü

SMR Metal Back Glenoid

ü ü

SMR TT Hybrid Glenoid

ü

SMR Cemented Glenoid

ü

SMR Cemented Glenoid 3-Pegs

ü

SMR Metal Back Liner

ü

SMR Connectors with screw

ü

SMR Glenospheres

ü

SMR TT Augmented 360 (Baseplate and Peg)

ü

SMR TT Hybrid Glenoid Reverse Baseplate + Screw

ü

SMR Bone Screw

A= Anatomic / R=Reverse

Table 2 SMR Shoulder System components compatible with PRIMA Humeral System

The following table identifies the allowed ( ü ) / not allowed ( x ) combinations between the SMR Lateralized Connectors and the SMR Glenospheres.

HP Glenospheres

Connectors

Diameter

Ecc.

LAT+4 LAT+2

ü ü

Ø40mm

0

x

ü

Ø44mm

0

x

x

Ø44mm Corrective

4

Note. SMR Humeral Heads diameter 38 mm are not allowed to be coupled to PRIMA Humeral System. SMR Humeral Heads Dia 40mm H13, Dia 42mm H13 both neutral and eccentrical versions are not allowed to be coupled to PRIMA Humeral System. Note. The CTA head must be used only in case of good stability of PRIMA Stem. If the CTA Head has to be used, an Eccentric Adaptor must be used (the coupling with the Concentric Adaptor is not allowed) and that the eccentricity is in the cranial direction only.