2.1 COMMON SPECIFICATIONS The common specifications for all FMS Series frames are described in the following sections. 2.1.1 POWER FMS frames feature a self-regulating power input module. There is no need to adjust your system with jumpers or voltage regulators based on the source input power. FMS frames are single phase voltage instruments capable of support source power at 100, 120, 220 and 240Vac ±10%. Frequency is 47 to 63 Hz. 2.1.2 FUSE RATING All FMS frames require two (2) Slo-Blo type fuse per the informationin the following table.
2.1.6 MEASUREMENT ACCURACIES The measurement accuracies for all FMS systems are shown in the table below.
PARAMETER
ACCURACY
FLC Sensor Load Accuracy FMS Frame Speed Accuracy FMS Frame Distance Accuracy
±0.1% Full Scale ±0.2% of set speed
Better than 0.02% Full Scale (No Load)
3.0 LIFTING AND HANDLING Your FMS force measurement system is packed in a sturdy plywood crate. The protective crate is environmentally friendly and assembled with clamps that secure the walls together without screws. Clamps may be removed with the special clamping tool supplied or using a screw driver. Your FMS system is intended to be used on a table or work bench. Make sure the table or work bench has sufficient strength to support the weight of the testing frame. Before you locate your frame, please ensure the following: • Read the section Site Considerations. Verify that all conditions are met prior to transporting and locating your frame. • Your equipment operators are trained and appropriately licenses (if necessary) to your local safety standards. • At the location where your frame is to be established, there is sufficient ceiling clearance from the top of the frame. • You have adequate protective materials to protect your frame when moving to its intended site. 3.1 GENERAL PRECAUTIONS WARNING Your FMS frame is heavy. Use at least two adults to lift the frame. CAUTION Be careful to not put any strain on the Emergency Stop or other controls located on the front panel of your system. Make sure the crosshead travel limits are clear and free from becoming entangled during lifting.
TABLE 2.1 - FMS FUSE RATING FMS MODEL
FUSE RATING
SIZE
QTY
FMS500 FMS1000 FMS2500 FMS5000
250V 3.15A Slo-Blo 5 x 20mm 250V 3.15A Slo-Blo 5 x 20mm 250V 3.15A Slo-Blo 5 x 20mm 250V 6.3A Slo-Blo 5 x 20mm
2 2 2 2
2.1.3 WEIGHT FMS frames are heavy due to their construction to minimize deflection and maximize stiffness. The weights shown in Table 2.1 are frame weights only. WARNING Starrett recommends that at least two (2) adults be used to lift an FMS frame. Do NOT use fork lift equipment to lift an FMS frame since the forks may damage the granite base.
TABLE 2.1 - FMS FRAME WEIGHT FMS MODEL
POUNDS (LBS)
KILOGRAMS (KG)
FMS500 FMS1000 FMS2500 FMS5000
135 lbs 170 lbs 195 lbs 195 lbs
61 kg 77 kg 88 kg 88 kg
2.1.4 NOISE The FMS frames are designed to limit noise produced by the frame to less than 70dB(A). 2.1.5 ENVIRONMENTAL The following environmental specifications are common to all FMS frames.
PARAMETER
SPECIFICATION +50° to +100°F +10° to +38°C -40° to +150°F -40° to +66°C
Operating Temperature
Storage Temperature
Humidity
+10% to +90%, non-condensing
Designed for use under normal laboratory conditions. Protective measures may be required if location is subject to excessive dust, corrosive fumes, electromagnetic fields or hazardous conditions.
Environmental Conditions
10 PKG08823 -UMFMS