FMM User Manual

ENGLISH

2.1 COMMON SPECIFICATIONS The common specifications for all FMM frames are described in the following sections. 2.1.1 POWER FMM frames feature a self-regulating power input module. There is no need to adjust your system with jumpers or voltage regulators based on the source input power. FMM frames are single phase voltage instruments capable of support source power at 100, 120, 220 and 240Vac ±10%. Frequency is 47 to 63 Hz. 2.1.2 NOISE The FMM frames are designed to limit noise produced by the frame to less than 70dB(A). 2.1.3 ENVIRONMENTAL The following environmental specifications are common to all FMM frames.

3.0 LIFTING AND HANDLING Your FMM force measurement system is packed in a sturdy plywood crate. The protective crate is environmentally friendly and assembled with clamps that secure the walls together without screws. Clamps may be removed with the special clamping tool supplied or using a screw driver. Your FMM system is intended to be used on a table or work bench. Make sure the table or work bench has sufficient strength to support the weight of the testing frame. Before you locate your frame, please ensure the following: Read the section Site Considerations. Verify that all conditions are met prior to transporting and locating your frame. Your equipment operators are trained and appropriately licenses (if necessary) to your local safety standards. At the location where your frame is to be established, there is sufficient ceiling clearance from the top of the frame. You have adequate protective materials to protect your frame when moving to its intended site. 3.1 GENERAL PRECAUTIONS WARNING At least two adults to should used to move the FMM frame. CAUTION Be careful to not put any strain on the Emergency Stop or other controls located on the front panel of your system. Make sure the crosshead travel limits are clear and free from becoming entangled during lifting. 3.2 UNPACKING YOUR SYSTEM The FMM test frame is shipped in a heavy-duty corrugated carton. Foam inserts are used on the base, top and mid-section to help dampen vibration during shipment. WARNING Use caution if opening corrugated carton with a knife or utility blade. It is recommended you un-staple the top portion of the corrugated carton from the bottom base. • Remove the stapes that hold the top carton to the bottom base. • Remove the top cap. Leave the foam insert in place. • Remove the foam inserts. • Remove the Accessory boxes. NOTE Save the carton in case you need to store or ship your FMM frame at a later date. 3.3 POSITIONING YOUR SYSTEM FOR USE Using at least two (2) adults, firmly grasp the frame base. Lift the frame off of the foam base. Place the frame on to its operating position.

PARAMETER

SPECIFICATION +40° to +110°F +5° to +43°C -40° to +150°F -40° to +66°C

Operating Temperature

Storage Temperature

Humidity

+10% to +90%, non-condensing

Designed for use under normal laboratory conditions. Protective measures may be required if location is subject to excessive dust, corrosive fumes, electromagnetic fields or hazardous conditions.

Environmental Conditions

2.1.4 MEASUREMENT ACCURACIES The measurement accuracies for all FMM systems are shown below.

PARAMETER

ACCURACY

DFC Force Gage Accuracy DFG Force Gage Accuracy

±0.1% Full Scale ±0.2% Full Scale ±0.1% Full Scale ±0.2% of set speed

BLC Sensor Accuracy

FMM Frame Speed Accuracy FMM Frame Distance Accuracy

Better than 0.02% Full Scale (No Load)

10 PKG08813-UMFMM