FMD & MMD User Manual

2.3.5 Cyclic Testing Cyclic testing is defined as tests where the frame constantly reverses direction. Cyclic testing on MMD and FMD test frames is limited to a continuous 24 hour duration @ 1Hz sampling. CAUTION The MMD and FMD test frames are not suitable for fatigue testing applications. 2.3.6 Load Hold Testing Constant load testing is defined as tests where the frame maintains a constant load for a specified period of time, e.g. creep and relaxation applications. Cyclic testing on MMD and FMD test frames is limited to a continuous 24 hour duration at 1Hz sampling rate.

2.3.7 Frame Weights MMD and FMD Series weights are listed below, including dimensions.

Machine Weight

Shipping Weight 1 ~320 kg ~700 lbs ~340 kg ~750 lbs ~360 kg ~800 lbs

Height mm/in

Width mm/in 787mm 31 in 787mm 31 in 787mm 31 in

Depth mm/in 736mm 29 in 736mm 29 in 736mm 29 in

Model

MMD-10K FMD-10K MMD-30K FMD-30K MMD-50K FMD-50K

136 kg 300 lbs 192 kg 425 lbs 225 kg 500 lbs

1626mm 64 in 1626mm 64 in 1626mm 64 in

Physical Weights and Dimensions

1 The shipping weight will vary depending on the accessories included with your order. Accessories are packaged in a single shipping crate. NOTE All shipping containers used for the 10K, 30K and 50K frames are 38.5” W x 36” D x 74” H (97.8 x 91.4 x 188.0 cm).

starrett.com

15

Made with FlippingBook - professional solution for displaying marketing and sales documents online