TECHNICAL SPECIFICATIONS
1cm 2
2cm 2
5cm 2
ULTRASOUND APPLICATOR FREQUENCY
240 VAC, da 1,0 a 0,42 A, 50/60 HZ
Main power supply
1MHz
3MHz
1MHz
3MHz
1MHz
3MHz
Effective Radiating Area (ERA) (cm²)
1
0.9
1.5
1
2.5
2.7
Operating mode Continuous Dimensions and weight See table Electrical safety class
Maximum output power in Continuous mode
2W 1.8W 3W 2W 5W 5.4W
Class II, Type BF
Maximum output power in Pulsed mode
4.5W 3W 7.5W 8.1W
3W 2.7W (*)
EN/IEC 606011, CAN/CSA-C22.2 No.601.1, UL Std. No 60601-1
Safety test
Maximum amplitude in Continuous mode
2W/cm 2
2W/cm 2
2W/cm 2
2W/cm 2
2W/cm 2
2W/cm 2
Medical Device Classification
IIA 93/42 60601-1
Maximum amplitude in Pulsed mode
3W/cm 2
3W/cm 2
3W/cm 2
3W/cm 2
3W/cm 2
3W/cm 2
ULTRASOUND SPECIFICATIONS
Frequency
1 MHz, 3 MHz
Cycles
10%, 20%, 50%, Continuous
WEIGHT (WITHOUT BATTERY)
DIMENSIONS AND WEIGHT
WIDTH
DEPTH’
HEIGHT
Pulse repetition 16, 48, 100 Hz
INTELECT MOBILE 2 BASE UNIT
1 – 31,25 ms - Max (ON): 31,25 ms - Min (OFF): 5 ms
Pulse duration
COMBO
34 cm
35.5 cm
15 cm
3.1 kg
ULTRASOUND
34 cm
35.5 cm
15 cm
2.8 kg
VACUUM SPECIFICATIONS
Input
20-25 VDC
STIM
25.5 cm
35.5 cm
15 cm
2.9 kg
Maximum peak current
TROLLEY SPECIFICATION
4 A
Trolley (supports a maximum load of 6.5 kg)
48 cm (max)
52 cm (max)
96 cm
10.1 kg
Vacuum adjustment
0-600 mbar +/- 5%
Trolley with vacuum
48 cm (max)
52 cm (max)
96 cm
11.5 kg
Vacuum mode
Continuous or pulsed
Trolley + base unit
-
-
111 cm
-
71