Fluke 50 Series II Thermometers Lab accuracy in a field thermometer
Fluke FoodPro™ Food Safety Thermometer Speciality Model The Fluke FoodPro Thermometer provides a complete temperature
The Fluke 50 Series II Contact Thermometers offer fast response and laboratory accuracy in a rugged handheld test tool • Laboratory accuracy: ±(005% + 03°C) • Large backlit dual display • Min/Max/Avg • Electronic offset function maximises overall accuracy • Supports a wide range of thermocouple types • Temperatures displayed in°C,°F, or Kelvin (K)
measurement and monitoring solution for foodservice professionals Designed to enable foodservice professionals to quickly and accurately measure food product temperatures and to help detect potential food safety problems The Fluke FoodPro non-contact thermometer is the first line of defense against improper receiving, storage and holding temperatures Using an infrared measurement technique, the highly accurate FoodPro quickly reads surface temperatures, allowing workers to make rapid and frequent temperature measurements—faster than contact units and without fear of cross contamination The easy- to-see illumination light shows you the measurement area
• Three year warranty Powerful data logging capabilities The Fluke 53 II B and 54 II B can log up to 500 points of data to internal memory • User-adjustable recording intervals • Real-time clock captures the exact event time • Recall function lets logged data be easily reviewed • Download data to optional FlukeView ® PC software
Fluke 54 II B
Fluke 53 II B
Fluke 51 II
Fluke 52 II
Specifications
51 II
52 II
53 II B
54 II B
Thermocouple types Number of inputs
J,K,T,E Single
J,K,T,E
J,K,T,E,N,R,S J,K,T,E,N,R,S
Dual
Single
Dual
Time stamp
Relative Time Relative Time Time of Day
Time of Day
Splash/Dust resistant
l
l
l
l
Dual display with backlight Min/Max/Avg recording (T 1 -T 2 ) True differential Data logging up to 500 points IR data port for interface to PC Compatible with FlukeView ® Software
l
l
l
l
l
l
l
l
- - -
-
l
l
- -
Included Accessories Fluke 50 Series Thermometers: Impact
l
l
l
l
-
-
l
l
Absorbing Holster, Two Bead Probe Thermocouples 80PK-1 (54 and 52 only), One Bead Probe Thermocouple, 80PK-1 (51 and 53 only) Fluke FoodPro (FP): AA Battery and Quick Start Guide Ordering Information FLUKE 51 II Single Input Digital Thermometer FLUKE 52 II Dual Input Digital Thermometer
Battery life
1000 hours typical, AA 173 mm × 86 mm × 38 mm
Size (HxWxD)
Weight
04 kg
Temperature range and accuracy J-type Thermocouples
-210°C to 1200°C -200°C to 1372°C -250°C to 400°C -150°C to 1000°C -200°C to 1300°C
K-type Thermocouples T-type Thermocouples E-type Thermocouples N-type** Thermocouples
R** and S-type** Thermocouples
0°C to 1767°C
Temperature accuracy above –100°C: J, K, T, E, and N-type**
FLUKE 53 II B
Single Input Digital Thermometer with Data Logging Dual Input Digital Thermometer with Data Logging
±[ 005% + 03°C ] ±[ 005% + 04°C ]
R** and S-type**
Temperature accuracy below –100°C: J, K, E, and N-types
FLUKE 54 II B
±[ 020% + 03°C ] ±[ 050% + 03°C ]
T-type
FLUKE FP FoodPro™ Food Safety Thermometer