Fluke 2024 IG Test Tools Catalogue

Fluke 725 and 726 Multifunction Process Calibrators

More calibration power! • Two separate channels; measure, source and view process signals simultaneously • Measure volts, mA, RTDs, thermocouples, frequency, and resistance to test sensors and transmitters • Source/simulate volts, mA, thermocouples, RTDs, frequency, and pressure to calibrate transmitters • Measure or source pressure using any of 50 Fluke xx Pressure Modules • Source mA with simultaneous pressure measurement to conduct valve and I/P tests • Perform fast linearity tests with auto step and auto ramp • Power transmitters during test using 24 V loop supply and simultaneous mA measurement • Store frequently-used test setups for later use • Three year warranty 726 additional features: • More precise measurement and calibration source performance, accuracies of 0.01% • Transmitter error% calculation • Memory storage for up to 8 calibration results • Frequency totaliser and frequency pulse train source mode • HART mode inserts 250 ohm resistor in mA measure and source for compatibility with HART instrumentation • Integrated pressure switch test • Custom RTD curves, add calibration constants for certified RTD probes for enhanced temperature measurement

Recommended Accessories

TL220 SureGrip™ Industrial

C125 Soft Case See page 176

Test Lead Set See page 167

80PK-27 Industrial Surface Temperature Probe See page 169

700PRV Pressure Relief Valve Kit See page 186

Fluke 726

Fluke 725

750PXX Pressure Module See page 137

Function or Source Range or Type

Resolution Accuracy

Notes

0.01 mV 0.001 V 0.001 V 0.001 V

0.01% 0.02% Rdg + 2 LSD

Max load, 1 mA

Voltage

0 to 100 mV 725: 0 to 10V (source) 0 to 20V (source) 725/726: 0 to 30V (measure)

mA

0 to 24

0.001 mA 0.01%; 0.02% Rdg +2 LSD

Max load, 725/726: 1000Ω

mV (TC terminals)

-10.00 mV to +75.00 mV 0.01 mV

0.01% 0.025% of range + 1 LSD

Ohms

15Ω to 3200Ω 5Ω to 4000Ω

0.01Ω to 0.1Ω 0.10Ω to 1.0Ω 0.015%

Hz - CPM

2.0 to 1000 CPM 1 to 1000 Hz 1.0 to 10.0 kHz 10.0 to 15.0 kHz 725/726: 24 V DC

0.1 CPM 1 Hz 0.1 kHz 0.1 kHz

±0.05% ±0.05% ±0.25% ±0.05%

Source; 5V p-p 1V - 20 V p-p squarewave, -0.1 V offset

Loop Supply

N/A

10%

Included Accessories TL75 Test Lead Set (TL75), AC70A Test Clip Set (725), AC72 Test Clip Set (726), Stackable Test Lead Set, Four AA Batteries, Replacement Fuse (725), User Manual on CD, NIST Traceable Calibration Certificate. Ordering Information FLUKE-725 Multifunction

T/C

J, K, T, E, L, N, U, XK

0.1ºC

to 0.7ºC to 0.2ºC to 1.7ºC to 1.2ºC

T/C

B, R, S, BP

1ºC

RTDs

Cu (10) , Ni 120 (672) Pt 100, 200, 500, 1000 (385)

0.01ºC

to 0.15ºC

Pt 100 (3916), Pt 100 (3926)

0.1ºC

to 0.2ºC

Simultaneous function capability

Channel A Channel B

24.000 mA DC

M M — M — — — —

M or S

24.000 mA DC with 24 V loop supply

100.00 mV DC

M or S

30.000 V DC measure

Process Calibrator

20.000 V DC measure; 10.000 V DC source; 20.000 V DC source

M or S M or S M or S M or S

FLUKE-726 Precision

15 to 3200 Ohms (725); 5 to 4000 Ω (726) Thermocouple J, K, T, E, R, S, B, M, L, U, N, XK, BP

Multifunction Process Calibrator

RTD Cu 10 , Ni120; Pt100 (392); Pt100 (JIS); Pt100, 200, 500, 1000 (385), XK and BP

Pressure (requires Fluke 750PXX Modules)

M M used as S

Frequency; 10 kHz; (15 kHz) M = Measure, S = Source/Simulate

M or S

C ontact your local distributor for price and availability. fluke.co/wtb

Unique 726 features are in bold

fluke.co/calibration

Process Calibration Tools

125

Made with FlippingBook Learn more on our blog