Fluke 154 HART ® Calibration Assistant
HART ® Calibration assistance at your fingertips. Combine with a Fluke Calibrator for a total HART Calibration Solution. The 154 puts HART device configuration at your fingertips. An Android™ based tablet user interface makes HART configuration easy. The external wireless modem allows you to connect the modem to the transmitter being tested so you can communicate to the device remotely. HART transmitters can often be found in confined spaces that are difficult to access, or are out of reach. The Fluke 154 helps reduce the need to stand right next to the device being tested or configured so you can work from a safer, more convenient location. When you combine the Fluke 154 Communicator with a Fluke 750 Series Documenting Process Calibrator, or a Fluke 720 Series Multifunction Process Calibrator you have a complete HART calibration and configuration solution. • Full HART Device Description (DD) support of all HART devices • Configure HART devices • Perform HART trim on HART devices when used with a Fluke 750 or 720 Series Calibrator • Monitor device PV, SV, TV, QV and other measured HART variables • Quarterly HART DD updates for free • Convenient wireless connectivity to HART modem • Easy to use, fast connect and view HART data • Store HART device configurations • Long range wireless communication up to 76.2 metres • Three year warranty
Fluke 154
Tablet Operating System
Android 5.0 or newer
Processor
ARM or Atom
Memory RAM Memory ROM
1 GB 2 GB
Screen
960 × 540 qHD
Communication Port HART ® Modem Enclosure Dimensions (LxWxH) Total Unit Weight HART Cable Length HART Cable Connectors
Bluetooth
10.16 cm × 2.5 cm × 6.9 cm
0.16 kg
1.2 m
Mini-Grabber
HART ® Interface Interface
Capacitive coupling
Included Accessories Tablet Charger and USB,
Output
600 mVpp
Leakage
< 10 uA
Micro-cable, Fluke HART ® App (installed), HART Modem, USB Mini-Cable, Hook Test Clips, Alligator Clips, Magnetic Hanging Strap, Case, Manual installed on Tablet. Ordering Information Fluke 154 HART Calibration Assistant
Connections Specifications
Across loop resistor or HART field device HART 4, HART 5, HART 6, HART 7
HART Physical Layer Spec
HCF_SPEC-54 Bluetooth v2.0
Specification
Frequency
2.4 GHz
Class 1 Master Distance
275 ft (83.8 m)
Battery Type Battery Life
Rechargeable Li-ion
19 hours continuous use
Humidity
0% to 90% (non-condensing)
C ontact your local distributor for price and availability. fluke.co/wtb
fluke.co/calibration
Process Calibration Tools
127