Fluke 717, 718 and 719 Pressure Calibrators Accurate and rugged, temperature calibration tools
Fluke 717 Pressure Calibrators • Measure up to 10,000 psi/690 bar sensor (10,000G model) • Compatible with non-corrosive gases and liquids
Specialty Model Fluke 718Ex Intrinsically Safe Pressure Calibrator The Fluke 718Ex offers a convenient, self contained solution for pressure measurements and calibration.
Fluke 718 Pressure Calibrators • 1, 30, 100 and 300 psi ranges available • Pump to 300 psi, 20 bar with internal pump (718-300G) • 718-1G includes special low volume pump and high measurement resolution for low pressure calibration Fluke 719 Electric Pressure Calibrators • Electric pump enables one-handed pressure pumping and makes pressure calibration quicker and easier • Best in class 0.025% pressure measurement accuracy • Programmable pump limit settings can eliminate over-pressurisation—set ‘pump to’ pressure values • Two ranges, 30 psi and 100 psi
• ATEX II 1 G Ex ia IIC T4 • 30, 100, and 300 psi ranges (2, 7, and 20 bar) • Pressure measurement to 0.05% of full span, using an internal pressure sensor. • Pressure measurement to 200 bar using any of eight intrinsically safe Fluke 750PEx Pressure Modules
Fluke 718
Fluke 719
Recommended Accessories
Fluke 717, 718 and 719 common features • Pressure switch test makes a difficult task easy, captures set, reset and deadband values of a pressure switch • 717 and 718 measure pressure to 0.025% of full scale with internal sensor • Measure mA with 0.015% accuracy and 0.001 mA resolution, while sourcing 24 V loop power • Three year warranty
700ILF In-Line Filter See page 186
700LTP-1 Low Pressure Test Pump See page 186
Fluke 717
C116 Large Soft Case See page 175
Models 717-1G 718-1G 717-15G
Range
Resolution
Comment
+0.0001 psi, 0.0001 kPa, 0.0001 in H20 0.001 ps1, 0.01 kPa, 0.01 in H20 0.001 psi, 0.001 kPa, 0.001 in H20
+/- 1 psi, 27.5 in H20, (-7 - +7 kPa) -12 to 15 psi, -83 to 103 kPa
Fluke 718 dry air only and includes on board pump
—
717-30G
Fluke 718 and Fluke 719 dry air only and includes on board pump Fluke 718 and Fluke 719 dry air only and includes on board pump Fluke 718 dry air only and includes on board pump Compatible with non-corrosive gases and fluids Compatible with non-corrosive gases and fluids Compatible with non-corrosive gases and fluids Compatible with non-corrosive gases and fluids Compatible with non-corrosive gases and fluids
Included Accessories Fluke 717/718/719: Protective Yellow Holster with Test Lead Storage, Instruction Sheet, 9 V Alkaline Battery (1x 717, 2x 718 & 719), Test Leads and Alligator Clips, Hose Kit (719), NIST- Traceable Calibration Report & Data. Fluke 718Ex: Protective Red Holster, TL75 Test Leads, AC72 Test Clips, User Manual (CD), Fluke 718Ex CCD, NIST- Traceable Calibration Report & Data. Ordering Information FLUKE-717 10,000G - 1G Pressure Calibrator, 10,000 PSI - 1 PSI (10 models)
-12 to + 30 psi, (-83 to 207 kPa)
718-30G* and 719-30G 717-100G
-12 to + 100 psi, (-83 to 690 kPa)
0.01 psi, 0.01 kPa, 0.01 in H20
718-100G* 717-300G
-12 to +300 psi, (-83 to +2070 kPa)
0.01 psi, 0.01 kPa, 0.01 in H20
718-300G* and 719-100G 717-500G
0-500 psi, 0-3450 kPa 0.01 psi, 0.1 kPa
717-1000G
0-1000 psi, 0-6900 kPa 0.01 psi, 0.1 kPa
717-1500G
0-1500 psi, 0-10342 kPa 0.1 psi, 1 kPa
717-3000G
0-3000 psi, 0-20700 kPa 0.1 psi, 1 kPa
717-5000G
0-5000 psi, 0-34500 kPa 0.1 psi, 1 kPa
717-10000G 0-10000 psi, 0-69000 kPa Compatible with non-corrosive gases and fluids Pressure accuracy 0.035% of range, 1 year. 0.025% for 6 months, 717 and 718-1G: 0.05% 6 months, 1 year, 718 and 717-300G 0.035% 6 months, 0.05% 1 year mA measurement 0-24 mA range 0.001 mA resolution Accuracy: ±0.015% + 1 count Loop power 24 V DC Accuracy: ±10% 1 psi, 1 kPa
FLUKE-718 300G, 100G, 30G FLUKE-718Ex 300G, 100G, 30G FLUKE-719 100G, 30G
Pressure Calibrator
Intrinsically Safe Pressure Calibrator Electric Pressure Calibrator 100PSI, 30 PSI
fluke.co/calibration
Process Calibration Tools
135
Made with FlippingBook Learn more on our blog