Fluke 730G Smart Digital Pressure Calibrator
Recommended Test Pumps
The Fluke 730G Smart Digital Pressure Calibrator with internal HART ® communication is designed in a pressure gauge form. In conjunction with a pressure test pump, the 730G enables high-efficiency and high-accuracy measurement and calibration of pressure gauges, pressure switches, pressure transmitters, which makes it ideal for mobile metrology labs, metrology institutes and enterprises to establish pressure laboratories. • Pressure measurement from ±10 in H2O/2.5 kPa to 10000 psi/69 MPa • Four absolute pressure measurement range: 15 PSIA (103 kPa), 30 PSIA (207 kPa), 100 PSIA (690 kPa), 300 PSIA (2 MPa) • Pressure measurement accuracy 0.05% of full scale, 0.02% optional and 0.01% current and voltage accuracy • Current measurement range: 0-24 mA; voltage measurement range: 0-30 VDC • Powerful HART communication, supports HART 7 • 24 V loop powered, no external power supply needed • Storage up to 128 sets of calibration data • Ingress protection: IP54 • 1-metre drop test • Three Year Warranty
Fluke 730LTP/40K • Pressure range -40 kPa to 40 kPa • Ultra-precision fine adjustment. The minimum controllable adjustment granularity is high
Fluke 730G
• Special insulation which minimises the influence of external temperature
Fluke 730LTP/600K • Pressure range -95 kPa to 600 kPa • Pressure is generated by the combination of the pressure rod and screw for faster and smoother pressure increase and decrease • Safely switch between pressure and vacuum
Fluke 730G
Specifications Current
0 to 24 mA 0 to 30 V DC
Voltage
Current range Voltage range
1 μA 1 mV
Accuracy 0.01% ±2 LSD all ranges @ 23°C ±5°C Temperature compensation From 20% FS/°C @ -10°C to +18°C and 28°C to 50°C Loop Compliance Voltage 24 V DC @ 20 mA
Temperature measurement accuracy inside the product
±0.5°C
Fluke 730PTP/10M • Pressure range -0.095 MPa to 10 MPa • Multi-stage boosting is used, which is 5 times the conventional efficiency • Safety lock ensures safe switching between positive and negative pressure
Drive Capability Size (HxWxD)
1200 Ω (without HART resistor), 950 Ω (with internal HART resistor)
155 mm × 130 mm × 44 mm
Weight
850 g
Fluke 730G/PCSW Software
The Fluke 730G/PCSW verification management software completes the calibration workflow by allowing for paperless management of pressure calibrations. Record calibration data, automate calibration tests and generate calibration reports with ease.
• Paperless testing and reporting: Enable paperless testing giving you the ability to automatically generate test reports and document results. • Manage tested equipment and pressure standards with ease. • Supported pressure devices: analog pressure gauges, digital pressure gauges, pressure transmitters, pressure sensors and pressure switches.
Fluke 730HTP/70M • Pressure range 0 MPa to 70 MPa • The pressure generation by the pressure rod can calibrate pressure meters with a large cavity • Fine adjustment via the screw has a high minimum controllable adjustment granularity
142
Process Calibration Tools
Pressure Calibration
Made with FlippingBook Learn more on our blog