Sterngold ERA® Product Catalog_666116 T

Sterngold | ERA® Attachment Systems

Common Causes of Wear

Common Causes of Wear for Attachments with Nylon Components

ERA ® FabricationTips for Plastic Females

The following suggestions from the dental laboratory technicians and dentists on our technical staff, as well as some of our field survey labs, will contribute to extending the life of the prosthesis and patient comfort. Patient Consistently Bites the Overdenture Into Place The patient should be instructed on how to insert the prosthesis into place with his/her fingers. Cleaning Abutments or Females with an Abrasive Cleaner Toothpaste can be very abrasive. The patient should be instructed to remove all traces of toothpaste after brushing. The abrasive in some toothpaste can cause extensive wear on the inside of the female when the male is inserted. Denture Cleaners Patients who use denture cleaner should be advised to follow the manufacturer’s instructions. Most require a soak time of only 5 to 15 minutes then rinse and store overnight in only water. Patient should avoid leaving the overdenture in the cleaning solution for extended time or overnight since many cleaners can have a detrimental effect on the surface of the nylon males and cause them to loose their retention. The cleaning solution can actually attack the nylon of the males, causing it to soften in approximately 1-2 months. Solutions containing Chlorine may cause the nylon males to become hard and brittle. This will cause premature wearing of the males, and eventually, some wear of the females. Insertion of the ERA ® Blue, Grey, Yellow, Red, Green or Purple ColoredMale The insertion of the blue, grey, yellow or red males prematurely can cause excessive wear in the male. Two additional retention males are available in the ERA-RV Partial version. They are green and purple.

Distal Extension (ReducedVertical andMicro) Although the ERA female is a plastic burnout pattern and can be cast in virtually any alloy, care should be taken to choose a hard alloy. The two most important aspects to consider are Vickers hardness and Ultimate Tensile strength. These attributes will provide optimum retention and long life of the female eyelet. A minimumVickers of 200 and an Ultimate Tensile of at least 75,000 psi is recommended. Overdenture We should pay particular attention to the cast Overdenture application of ERA. Prescriptions in these cases often request a yellow crown and bridge alloy. In choosing this kind of alloy, use a type IV metal that allows you to harden the finished casting to a high Vickers Hardness. There are many of these on the market today including Stern’s Apollo, a 46% type III that can be hardened to 230 Vickers. Hardening techniques for C+B metals are relatively standardized. First, anneal the finished casting by heating and quenching and then harden by heating and slow cooling. (Refer to specific alloy manufacturer for temperatures and detailed hardening technique.) This is a simple step that can ensure a successful case. WARNING: Great care must be taken when using debubblizers on plastic patterns. These materials may present problems in investing and burnout of the patterns. You must be sure that there are no puddles in or around the attachment. Also take care to fill the inside of the female pattern with a small brush before investing the rest of the pattern. Finishing the Cast ERA ® Females Females are divested in a normal fashion by either sandblasting lightly or stripping in ultra-sonic solution. Care should be taken that the inside of the female is not over sandblasted as this will oversize the female eyelet. Glass beads or light polishing with rubber points can be used to create a clean finished look to this interior surface. The outside surface is not a crucial dimension and can be rubber wheeled and polished but should not require stone or carbide finishing. Parallelism It’s ideal to have the attachment parallel, however, the attachment will function normally up to 8° out of parallel with the case’s path of insertion. If the attachments are off by more than 8° you might not realize maximum life span of the males. Path of Insertion We would like to stress the importance of a short flange in the anterior region. Appliances designed to engage the labial undercut will interfere with proper seating. In cases of deep labial undercut, it should be blocked out before the processing of the denture base. Engage no more than 1 mm of undercut. Full extension of the flange into the vestibule will still provide lip support and help prevent food entrapment.

Keep in mind that the least amount of retention required by the patient is best.

The dentist should always process the prosthesis first with the Black Fabrication Male, core out that male, and then snap in theWhite Male Attachment which is the least retentive. Then, if the patient wants more retention, the Orange Male should be used. Only step up one degree of retention at a time. Tobacco Chewing Silica in the juices created by chewing tobacco can be very abrasive, and may be a cause of wear to the male and the female. Pickup of Attachments Chairside The dentist should make sure that the patient does not bite down once the prosthesis has been placed. We cannot determine the strength of a patient’s bite. Too much pressure can cause the tissue to be displaced. Once the composite around the male has set and the pressure relaxed, the tissue returns to a normal state raising the attachment slightly. This can cause the attachment to snap in and out constantly as the patient talks or chews, causing excessive wear of the male. To avoid this problem, have the patient bite passively. As patient relaxes bite, the dentist should passively hold the prosthesis in place until the acrylic or composite cures.

13

ERA® PRODUCT CATALOG

WWW.STERNGOLD.COM

Made with FlippingBook - professional solution for displaying marketing and sales documents online