KMR Mx200 User Guide

2 Environmental and Safety Information

This section provides the environmental and safety information for the KMR system.

2.1 Environmental Considerations

The KMR system is factory calibrated under the standard laboratory environmental conditions shown in the following table.

Calibration Environment

Specification

Requirement

Ambient Temperature

20°C ± 0.5°C (68°F ± 5°F)

Humidity

40-60% RH

Temperature rate of change

0.5°C (1°F) per hour

Operating Environment

KMR system can be safely operated under the following environmental conditions:

Environmental Condition

Operational Requirement

Ambient Temperature

20°C ± 3°C (68°F ± 5°F)

Humidity

< 90% RH

If the system is to be operated under environmental conditions that are substantially different from those in the previous table, the system should be recalibrated under the expected conditions. You should also consider the material characteristics of the parts under inspection, for example, coefficients of thermal expansion. Numerical compensation might be required when measuring parts under conditions different from those controlling the stated dimensional specifications for these parts. Before performing critical parts measurements, allow the system to warm up and temperature stabilize for 30 minutes.

2.2 Safety Considerations

General Safety

The KMR system is designed for safety and proper ergonomics during normal use. Exercise caution when handling or moving the system to maintain calibration and measurement performance. Disconnect all power sources prior to moving or working on the equipment. Consult SKE if you have any questions regarding transporting, using, or maintaining the system.

Electrical Safety

The KMR system does not contain hazardous AC line voltages, as these are contained on the input side of the system’s AC adapters, which are UL listed. Even at these low voltages,