Mobile Phase Cleaner for UHPLC/HPLC
GLC Suction Filter 2 HPLC Suction filter with mobile phase cleaning function
Features
Removes contaminants from the mobile phase.
Does not affect gradient delay* to install this product in front o fthe pump (see Fig. 2).
Available for LCMS due to low bleed.
(*The delay form the time a gradient begins to be produced in the pump to when it
reaches the column inlet).
Flow
The GLC Suction Filter 2 is packed with activated carbon which acts as a chemcal trap and removes contaminants present in the mobile phase. This product is easy to install and user-replaceable.
GLC Suction Filter 2
Column
Detector
Activated carbon
Pump
Note: Connect 3.0mm outer diameter suction tube when you use this product.
Degasser
Fig. 1. Structure of the GLC Suction Filter 2 i. 1. Structure o
Fig. 2. Attachment position
Removal of contaminants in mobile phase (LC/UV) e.g. Removal of contaminants caused due to the storage of the mobile phase Ghost peaks are contaminant peaks that appear even when no sample is injected. One of the causes for ghost peaks is contamination in mobile phase due to improper storage or from external sources. We prepared contaminated ultrapure water that was stored in a plastic container (polyethylene) at room temperature for two weeks. Ghost peaks were observed when using the contaminated water as mobile phase. After installing the GLC Suction Filter 2, the previously observed ghost peaks were eliminated.
Normal Suction Filter GLC Suction Filter 2
/ 6 : ; , ( 2:
Analytical Conditions Instruments : Nexera XR Column
: Shim-pack VP-ODS (150 x 4.6 mm I.D.) Mobile phase: A ; Contaminated ultrapure water B ; Acetonitrile Time prog. : B Conc. 20% (0-5 min.) →
95% (5-25 min.) → 20% (25.01-40 min.)
Flow rate
: 1 mL /min.
Product Informatio n
Please read the manual carefully and condition this product before use. * The suction tube is not included. * When using ion-pair reagents in the mobile phase, the ion-pair reagents may be adsorbed to this product and may affect the retention time and peak shape. * Bleeding from this product may be seen. * This product does not remove all contaminants. * When the activated carbon reaches breakthrough, "ghost peaks" derived from mobile phase contaminants may be detected. In this case, replace the product. * The recommended replacement period is after passing 60L of mobile phase or 1 year after use.
Connecting tube: 3.0mm outer diameter suction tube Internal volume: 0.7 mL Pore size:10µm
GLC Suction Filter 2
1
370-02010-01
Pg A-122