Consumables Digital Catalogue

Column Chemistries Combining highly efficient core shell particle technology with a wide range of surface chemistries provides you with the best opportunity for optimal resolution. With different chemistry characteristics, Shim-pack Velox columns are suitable for use in a wide variety of applications and challenging separations.

SP-C18

C18

Biphenyl

PFPP

HILIC

USP Classification

L1

L1

L11

L43

L3

Sterically protected C18

Pentafluorophenyl propyl

Ligand type

C18

Biphenyl

None

Particle Size (µm)

1.8, 2.7, 5

1.8, 2.7, 5

1.8, 2.7, 5

1.8, 2.7, 5

2.7

Pore size (Å)

90

90

90

90

90

125 m 2 /g 130 m 2 /g 100 m 2 /g

125 m 2 /g 130 m 2 /g 100 m 2 /g

125 m 2 /g 130 m 2 /g 100 m 2 /g

125 m 2 /g 130 m 2 /g 100 m 2 /g

-

1.8 µm 2.7 µm

130 m 2 /g

Suraface Area

- -

5 µm

1.8 µm 2.7 µm

7 % 7 % 5 %

9 % 7 % 5 % YES

7 % 7 % 5 % YES

4 % 4 % 3 %

Carbon Load

NA

5 µm

-

End-Cap

No

No

No

pH range

1.0-8.0

2.0-8.0

1.5-8.0

2.0-8.0

2.0-8.0

1.8 µm 2.7 µm

100 MPa*

100 MPa*

100 MPa*

100 MPa*

-

Max Pressure

60 MPa 40 MPa

60 MPa 40 MPa

60 MPa 40 MPa

60 MPa 40 MPa

60 MPa

5 µm

-

* For maximum lifetime, recommended maximum pressure for 1.8 µm particles is 80MPa.

Lot to Lot Reproducibility We understand that lot to lot consistent performance of columns is required to maximize your laboratory performance. From one lot to the next, every Shim-pack Velox column you purchase will perform consistently.

Compounds (50 ng/mL) 1. Cortisol 2. 11-Deoxycortisol 3. Estradiol 4. Boldenone 5. Testosterone 6. Androstenedione 7. Progesterone

Conditions

Column

: Shim-pack Velox Biphenyl 2.7 µm, 3.0×100 mm (P/N: 227-32016-03 )

Flow Rate

: 0.7 mL/min

Column Temp. : 30 °C Sample Dilulent : Initial mobile phase Injection Volume : 5 µL Mobile Phase A

: 0.1 % formic acid in water

Mobile Phase B

: 0.1 % formic acid in acetonitrile Gradient: 40 % B (0 min) 80 % B (3 min)

Pg A-9

Made with FlippingBook Digital Proposal Creator