GC Columns
Capillary Columns General-Purpose Columns
SH-2560 Highly polar phase; biscyanopropyl polysiloxane—not bonded
Stationary phase selectivity optimized for isomer separation to ensure accurate quantification of critical cis/trans FAMEs.
Application-specific QC test guarantees consistent, reliable performance for AOAC 996.06 and AOCS Ce 1j-07 methods.
Excellent sample capacity; no peak distortion means easy, accurate peak integration.
Similar Phase(s): HP-88, CP-Sil 88, SP-2560, BPX-90, MEGA-10
ID
df
Temp. Range
100 m
0.25 mm 0.20 µm
20 to 250 °C
227-36311-01
Unsaturated FAMES QC Test Mix on SH-2560
Conc.
Structural
Conditions
Peaks
tR (min)
(µg/mL)
Nomenclature
Column
: SH-2560, 2560, 100 m, 0.25mm ID, 0.20 μm (227-36311-01) : Unsaturated FAMES QC test mix
1. Methyl octadecanoate
11.68 12.64 13.09 13.84 16.66 18.85 19.20 20.35 25.17 28.79 31.49
1.1 0.8 0.6 1.1 1.1 0.6 0.8 1.1 1.1 0.6 1.1
C18:0
2. Methyl elaidate 3. Methyl oleate
C18:1 (t9) C18:1 (c9)
Sample Diluent Injection
: Hexane
4. Methyl nonadecanoate 5. Methyl eicosanoate 6. Methyl eicosenoate 7. Methyl linolenate 8. Methyl heneicosanoate 9. Methyl docosanoate 10. Methyl eicosatrienoate 11. Methyl tricosanoate
C19:0 C20:0
: 1 µL split (split ratio 200:1) : Premium 4 mm Precision liner w/wool (cat.# 23305.1)
Liner
C20:1 (c11)
C18:3 (c9,c12,c15)
C21:0 C22:0
Inj. Temp
: 250 °C
Oven Temp Carrier Gas Flow Rate Detector
: 180 °C (hold 40 min) : H 2 , constant flow
C20:3 (c11,c14,c17)
C23:0
: 2.5 mL/min : FID, 275 °C
Make-up Gas Type : N 2 Hydrogen flow
: 40 mL/min : 400 mL/min
Air flow
1
Data Rate Instrument
: 50 Hz
: Agilent 7890B GC
4
5
2
8
3
9
7
11
6
10
10
12
14
16
18
20
22
24
26
28
30
32
34
Time (min)