Consumables Digital Catalogue

SH-CLP / SH-CLP II Dedicated columns for organochlorine pesticides and herbicides.

Low bleed - ideal for high-sensitivity GC-ECD or GCMS analyses.

Baseline separations in less than 10 minutes.

Analyze EPA Method 8081B, 8082A, 8151A, 504.1, 515, 508.1, and 552.2 compounds without time-consuming column changes.

Similar phases: DB-CLP1 / DB-CLP2

SH-CLP

ID

df

Temp. Range

30 m

0.32 µm

-20 to 320/340 °C

227-36266-01

0.32 mm

0.50 µm

-20 to 320/340 °C

221-75879-30

SH-CLP II

ID

df

Temp. Range

30 m

0.32 mm 0.25 µm

-20 to 320/340 °C

227-36267-01

Organochlorine Pesticides (US EPA Method 8081)

Peaks 1. 2,4,5,6-tetrachloro-m-xylene

2. α -BHC ( α -HCH) 3. γ -BHC (lindane) 4. β -BHC ( β -HCH) 5. δ -BHC ( δ -HCH) 6. heptachlor 7. aldrin 8. heptachlor epoxide

9. γ -chlordane 10. α -chlordane 11. 4,4'-DDE 12. endosulfan I

13. dieldrin 14. endrin 15. 4,4'-DDD 16. endosulfan II 17. 4,4'-DDT 18. endrin aldehyde 19. methoxychlor 20. endosulfan sulfate 21. endrin ketone 22. decachlorobiphenyl

Conditions

SH-CLP2, 30 m, 0.32 mm ID, 0.25 μm (P/N: 227-36267-01)

Column

: SH-CLP, 30 m, 0.32 mm ID, 0.50 μm (P/N: 221-75879-30)

Oven Temp

: 120 °C (hold 1 min) to 300 °C (hold 10 min) at 9 °C/min

Inj.

: Direct : 200 °C

Inj. Temp Detector

: ECD, 300 °C with anode purge

Dead time : 1.9 min Head pressure : 8.7 psi (constant) Flow rate

: 1.3 mL/min at 120 °C, He

Pg C-31

Made with FlippingBook Digital Proposal Creator