GC Columns
Capillary Columns Dedicated Columns
SH-BAC Plus 1 / SH-BAC Plus 2 Optimized column selectivities guarantee resolution of ethanol, internal standards, and frequently encountered interferences.
Robust and reproducible column chemistry ensures longer column lifetime and consistent results.
SH-BAC Plus 1
ID
df
Temp. Range
30 m
0.32 mm 1.80 µm
-20 to 240/260 °C
227-36260-01
0.53 mm 3.00 µm
-20 to 240/260 °C
227-36261-01
SH-BAC Plus 2 ID
df
Temp. Range
30 m
0.32 mm 0.60 µm
-20 to 240/260 °C
227-36263-01
0.53 mm 1.00 µm
-20 to 240/260 °C
227-36264-01
Analysis of Alcohol Compounds in Blood
Conditions
SH-BAC Plus 2
Instrument Headspace
: GC-2010 Plus AF + HS-20
: Oven Temp.: 85 °C Vial Warming Time: 15 min Vial Pressurization Time: 1 min Injection Time: 0.5 min Sample Line Temp: 150 °C Vial Volume: 20 mL Vial Agitation: Off Vial Pressurization: 100 kPa Load Time: 0.5 min
Peaks 1. Acetaldehyde 2. Methanol 3. Ethanol 4. Acetone 5. Isopropanol 6. t-Butanol 7. 1-Propanol
Needle Flash Time: 0.5 min Transfer Line Temp: 150 °C
Column
: SH-BAC Plus 2, 30 m, 0.32 mm ID, 0.60 μm (P/N: 227-36263-01) SH-BAC Plus 1, 30 m, 0.32 mm ID, 1.80 μm (P/N: 227-36260-01)
SH-BAC Plus 1
Column Temp : 40 °C Inj.
Peaks 1. Acetaldehyde 2. Methanol 3. Ethanol 4. Acetone 5. Isopropanol 6. t-Butanol 7. 1-Propanol
: Split (split ratio: 20:1)
Carrier Gas
: He, 100 kPa
Detector : FID, 250 °C Makeup Gas : He, 30 mL/min Hydrogen : 40 mL/min Air : 400 mL/min
Pg C-34