Consumables Digital Catalogue

Capillary Columns PLOT Columns

SH-Msieve 5A Stationary phase: Molecular sieve 5A

Easily separate permanent gases at temperatures above ambient.

Improve accuracy with sharp, symmetrical peaks for argon, oxygen, and carbon monoxide.

Similar phases: HP-PLOT Molesieve, CP-Molsieve 5A, Molsieve 5A PLOT

ID

df

Temp. Range

30 m

0.32 mm

30 µm

to 300 °C

227-36611-02

0.53 mm

50 µm

to 300 °C

221-75763-30

Analysis of Inorganic Gas

Ar m/z 40

O 2 m/z 32

N 2 m/z 28

CH 4 m/z 16

CO m/z 28

Peaks

1. Ar 2. CO 3. O 2 4. N 2 5. CH 4

5.65 ppm 5.44 ppm 5.54 ppm 5.40 ppm 5.71 ppm

Conditions

Instrument

: GCMS-QP2010 Ultra

Column : SH-Msieve 5A, 30 m, 0.32 mm ID, 30 μm (P/N: 227-36611-02) Sample injection : Gas sampler (1 mL loop volume) (P/N: 223-57653-91) Inj. Mode : Split (split ratio: 50:1) Inj. Temp : 200 °C Control Mode : Pressure (100 kPa) Carrier Gas : Helium Oven Temp

Ar

CH 4

N 2

O 2

: 35 °C (hold 2 min) to 150°C at 10°C/min (hold 5 min)

CO

Detector

: MS

Interface Temp: 200 °C Ion Source Temp: 200 °C Measurement Mode: Scan (m/z 10 to100) Event Time: 0.5 sec Ionization Method: EI Emission Current:150 μA

Pg C-41