Consumables Digital Catalogue

Rt-XLSulfur Columns (Packed and Micropacked) Rt-XLSulfur packed and micropacked columns are designed for ppb-level sulfur analysis. Every component of the sample pathway is treated to provide the highest degree of inertness for reactive, low-level sulfur compounds. The porous polymer phase features a unique surface modification, which results in excellent peak symmetry and thermal stability to 300 °C. The high performance and reproducibility of the Rt-XLSulfur column enables resolution and quantitation of COS, H 2 S, SO 2 , CH 3 SH, (CH 3 ) 2 S 2 at low ppb concentrations. These sulfur compounds typically are found in pulp mill byproducts, natural gas, and petroleum products.

Features:

Optimized columns for low pbbv sulfur analyses.

Eliminate the need for PTFE tubing.

Column and end-fittings are Sulfinert® treated for maximum inertness.

Figure 1 The Rt-XLSulfur column analyzes 50 ppb levels of sulfur compounds, providing low bleed and good symmetry.

Peaks 1. Hydrogen sulfide 2. Carbonyl sulfide 3. Methyl mercaptan

4. Ethyl mercaptan 5. Dimethyl sulfide 6. Dimethyl disulfide

Conditions

Column : Rt-XLSulfur, 1 m, 0.75 mm ID Sample conc. : 1 mL of 50ppbv each sulfur compound Injection : sample valve Oven Temp. : 60 °C to 230 °C at 15 °C/min. Carrier Gas : He, constant flow Flow Rate: : 9 mL/min. Detector : SCD

Column

OD

ID / mm

Length / m

1.0 2.0 1.0 2.0 1.0 2.0 1.0 2.0

1/8”

2.0

Rt-XLSulfur Columns (packed)

1/16”

3.2

1/16”

1.0

Rt-XLSulfur Columns (micropacked)

0.95mm

0.75mm

For availability and ordering information on custom columns, please contact your representative direct sales/distributors.

Pg C-55

Made with FlippingBook Digital Proposal Creator