HuFriedyGroup | DentaPure Cleaning Shocking EN

CLEANING/SHOCKING DENTAL UNIT WATERLINES

Dental Water Management

When a DentaPure™ Bottle Cartridge is Installed (DPI365B)

Dental unit waterlines must not be cleaned/ shocked with a DentaPure™ Cartridge in place. If this happens, use iodine test strips to determine that the level of iodine is above 0.5 ppm, or replace the cartridge.

1. Remove Perform a two-minute flush

4. Once the cleaning/ shocking process is complete: • Remove the non-functional demo unit from the luer fitting on the pick-up tube. • If installing a new DentaPure™ Cartridge, follow the included Instructions for Use. • If reusing the DentaPure™ Cartridge that was removed prior to cleaning/socking, remove it from the zip-top bag and re-install on the pick- up tube. • Ensure that you are re-installing the functional cartridge by looking for the presence of the filter in the end cap. • Operate dental unit instruments until all air is purged from the water system. Resume system use.

• Remove the DentaPure™ Cartridge from the luer fitting by turning it clockwise. The luer fitting should remain attached to the pick-up tube. • If installing a new DentaPure™ Cartridge, discard the depleted DentaPure™ Cartridge in your regular office trash. • If continuing use of the removed cartridge, place it in a zip-top bag labeled with the operatory number and initial cartridge installation date (estimated date is acceptable). DO NOT DISCARD. 2. Install • Install the non-functional DentaPure™ Cartridge demo unit onto the luer fitting on the pick-up tube. • Take note of the “Non-Functional Demo Unit Only. Not for Use in Dental Facilities” etching on the demo unit and the absence of a filter inside the end cap. 3. Clean/shock your DUWLs • Follow the Instructions for Use (IFU) that came with your cleaning/shocking product.

If you plan to use the non-functional demo unit in the future: • Thoroughly rinse the inside and outside with warm water. • Shake out excess water and allow to dry completely. • Place the dry non-functional cartridge in a zip-top bag with a copy of the Instructions for Use and label the bag: Non-functional DentaPure™ Cartridge Demo Unit for cleaning/shocking.

Visit us online at HuFriedyGroup.eu

Free Call: 00800 48 37 43 39 | Free Fax: 00800 48 37 43 40 E-Mail: info@hu-friedy.eu | Website: HuFriedyGroup.eu @HuFriedyGroupEurope

Hu-Friedy Mfg. Co., LLC. • European Headquarters • Lyoner Str. 9 • 60528 Frankfurt am Main, Germany • HuFriedyGroup.eu All company and product names are trademarks of Hu-Friedy Mfg. Co. LLC., its affiliates or related companies, unless otherwise noted. ©2022 Hu-Friedy Mfg. Co., LLC. All rights reserved. HFL-729GB/ 1 022

Page 1

www.hufriedygroup.eu

Made with FlippingBook Annual report