L3 Performing a Test

4.3.2.1 Using Omit Exception - No You specify whether or not to omit an exception in the Pre Test Step of your test setup. The Omit Exception default is Yes. When Omit Exception is set to No, if an exception occurs, the test run will automatically end at the exception. The data for the test is available for analysis.

4.3.2.2 Using Omit Exception - Yes When the Omit Exception option is set to Yes, if an exception occurs, the test run will automatically end. A message is displayed indicating that the test run was ended due to an exception. Any results in the Results view are depicted with a red x-out over the entire results area. Additionally, the run in the Runs List is shown in red with a red strike-thru indicated an aborted test caused by an exception. When Omit Exception = Yes, you are omitting the data from this run, e.g. no data is collected for analysis. This test run is voided.

Pre Test Omit Exception = No If an exception is reached, save data for analysis

With this test setup, if a test run should measure a load of 10N BEFORE reaching the Target Distance of 100mm, the test will automatically stop. The data acquired up to the 10N will be saved for analysis because Omit Exception = No, e.g. do not omit exception data.

Exception Message When Pre Test Omit Exception = Yes, a message appears and test run stops

Exception Limit Stops test automatically if Load gets to 100N

16