4.4.2 Stop Test Control During a test run, you may stop the run by selecting the Start/Stop push button or the STOP TEST target on the Controller view. When a test run is stopped, it completes the run. You cannot resume a stopped test . NOTE Pressing the Stop function ends the current test run.
4.4.3 Pause Test Control When an extensometer is specified in your Sample Definition step, you have the ability to “pause” a test run. This “pause” allows you to remove the extensometer from the sample prior to a break in order to protect the extensometer from damage. Pressing the STOP TEST target or the Start/Stop button will “pause” the test run. To re-start the test run, press the PLAY target on the STOP target or press the Start/ Stop push button on your test frame.
NOTE A stopped test run cannot be resumed.
Press the STOP TEST target or push button to PAUSE the test when using an extensometer.
Shows an active test run. Press the Pause symbol to PAUSE the test run.
Stop Target Press to stop a test run
Press the STOP TEST target or push button to Resume a “paused test” when using an extensometer.
Shows a PAUSED test. Press arrow to Resume.
Stop /Pause/Play Target Press to stop to Pause and to Re-Start a test run when an extensometer is used