4.2.2 Load Bargraph The Controller view displays the current applied load being measured by the load cell sensor. The load is displayed from the center of the bargraph. When the bargraph is filled from the center upward, tensile load is being applied. When the bargraph is filled form the center downward, compressive load is being applied. The bargraph displays green when the measured load is within 80% of the load cell sensor’s full scale capacity. The bargraph displays yellow when the measured load is within 81% - 95% of the load cell sensor’s full scale capacity. The bargraph displays red when the measured load exceeds 95% of the load cell sensor’s full scale capacity. Color indication is used to represent the measured load versus the load sensor’s capacity as follows:
Bargraph Color Load Indication GREEN
The sensor is operating within its normal range 0 to 80% of full scale capacity. The sensor is operating in its caution range 81 to 95% of full scale capacity. The sensor is operating in its high range 96% and above. Be alert of the measured load overloading and potentially damaging the sensor. Load Sensor Status Indication CAUTION When the bargraph displays in red, it indicates that the load cell is nearing its full scale capacity and that it is near an overload condition. USE EXTREME CAUTION NOT TO EXCEED THE LOAD CELL’S FULL SCALE CAPACITY.
YELLOW
RED
Normal Operating Run Bargraph displays green- load cell sensor in normal range
Caution Operating Run Bargraph displays yellow- load cell sensor in caution range
Warning Operating Run Bargraph displays red- load cell sensor in warning range