5.5 Using the Point Tool The Point tool lets you measure any point along your graph trace. The graph trace for a Run is comprised of points based on the Sampling Rate you have specified for your test setup. The Point tool is used in Load-Distance modes. And the Point tool may be scoped along either the y-axis (Load) or on the x-axis (Distance, Time). All Point measurements have a specified Scope. The scope allows you to further define where a point is measured. Using the Scope units, you may define the scope for your point based on load, distance or time. For example, if you want to measure a point that is 100mS after the maximum load on your graph, you can set the maximum load coef- ficient ( L max) as an anchor for your point, and then set the scope at 0.1 seconds after the anchor. This will give you a result for load that is 100mS after your maximum load result.
Slope Intersect Point
Limit Point
Offset Yield Point
Point Tool types
Active coefficient is LPt.
Limit point algorithm being used to find the Distance (DPt) at a load limit target of 50N.
Load Limit Point Data Definition Menu Showing result at 50N Load Limit