L3 System Settings

8.9.6 Disable Overload The Disable Overload function is intended for use during load cell calibration, where you need to disable overload protection. The Disable Overload function is password protected and may be accessed by authorized Starrett personnel. Contact your Starrett representative if you need to access the Disable Overload function. WARNING

8.10 Tests Setting The Tests setting is used to specify how certain Test functions perform. These settings include the ability of the user to edit a locked test; the use of prefixes in the test name; test numbering, and how tests appear on the main Home Menu.

The Disable Overload function removes all load cell sensor protection. The Disable Overload function is used during load cell sensor calibration. Disable Overload should NEVER be disabled during normal system use, e.g. all use except for load cell sensor calibration. Disabling overload protection may lead to permanent damage to a connected load cell sensor and potential permanent damage to your test frame.

L3 TESTS Setting Shows options associated with test setups

L3 LOAD Setting Shows required Password entry dialog to Disable Overload protection