Fluke Australia 2020 Test Tools Catalogue

Fluke 787B and 789 ProcessMeter ™

Recommended Accessories

Fluke 787B and 789 key features • 20 mA DC current source/measure/simulate • Simultaneous mA and % of scale readout • DMM designed to meet 1000 volt IEC 61010 CAT III and 600 V CAT IV standards • Fluke Connect ® compatibility for wireless data logging (with IR3000FC module) • Precision 1000 V, 440 mA true-RMS digital multimeter • Frequency measurement to 20 kHz • Min/Max/Average/Hold/Relative modes • Diode test and continuity beeper • Manual step (100 %, 25 %, Coarse, Fine) plus Auto Step and Auto Ramp • Externally accessible battery for easy battery changes • Three year warranty Fluke 789 additional features • 24 V loop power supply • HART mode setting with loop power (adds 250 ohm resistor) troubleshooting capabilities by combining the power of a safety rated digital multimeter and mA loop calibrator into a single, compact test tool. Whether you only need to source and simulate mA, or need a 24V loop power supply, Fluke has a ProcessMeter™ test tool designed specifically to meet your needs. Double the troubleshooting power so you can do more, while carrying a lot less The Fluke 787B and 789 ProcessMeter™ double

IR3000 FC Fluke Connect ® ir3000 Connector See page 155

PV350 Pressure Vacuum Module See page 155

TL220 SureGrip™ Industrial

80T-150UA Universal Temperature Probe See page 146

Test Lead Set See page 144

Fluke 789

Fluke 787B

80PK-8 Pipe Clamp Temperature Probe See page 146

TPAK ToolPak Meter Hanging Kit See page 154

Recommended Kits Buy more, Save more

Best Accuracy (% of reading + LSD)

Measurement Function

Range and Resolution

V DC

400.0 mV, 4.000 V, 40.00 V, 400.0 V, 1000 V 400.0 mV, 4.000 V, 40.00 V, 400.0 V, 1000 V

0.1 % + 1

V AC (True-RMS)

0.7 % + 2

mA DC

30.000 mA

.05 % + 2 0.2 % + 2

Fluke Connect ® 789 ProcessMeter™ and Temperature Kit

A DC A AC

1.000 A (0.440 A continuous) 1.000 A (0.440 A continuous) 400.0 Ohms, 4.000 k, 40.00 k, 400.0 k, 4.0 M, 40 M

1 % + 2

For more details see page 172

Resistance

0.2 % + 1

Frequency (0.5 Hz to 20 kHz)

199.99 Hz, 1999.9 Hz, 19.999 kHz .005 % + 1

Diode Rest

2.000 V (shows diode voltage drop) Beeps for resistance < approx. 100 ohms

2 % + 1

Included Accessories TL71 Test Lead Set, AC72 Alligator Clips, four AA Alkaline Batteries (installed), User Manual and Quick Reference Guide.

Continuity

Output Function Range and Resolution

Drive Capability Accuracy

Ordering Information

DC Current Output (Internal Battery Operation) DC Current Simulate (Ext. 15 V to 48 V loop supply)

0.000 to 20.000 mA or 4.000 to 20.000 mA, (selectable at power-up) Over-range to 24.000 mA 0.000 to 20.000 mA or 4.000 to 20.000 mA, (selectable at power-up) Over-range to 24.000 mA

24 V compliance or, 1,200 ohms, @ 20 mA

.05 %

ProcessMeter ™ ProcessMeter ™

FLUKE-789 FLUKE-787B FLUKE-789/ T3000FC

1000 ohms, @ 20 mA

.05 %

Fluke Connect 789 ProcessMeter ™ & Temperature Kit Fluke Connect Infrared Connector FlukeView Forms Software (incl. Interface Cable)

FLUKE-IR3000 FC

24 V Loop Supply* Minimum 24 V

250 ohms @ 20 mA > 24 V

Current Adjustment Modes

Manual: Coarse, Fine, 25 % and 100 % step Automatic: Slow Ramp, Fast Ramp, 25 % step

FVF-SC2

*24 V Loop Supply available on 789 ProcessMeter TM only.

C ontact your local distributor for price and availability. www.fluke.com/au/wtb

106

Process Calibration Tools

Made with FlippingBook Learn more on our blog