Reference Section • Cassettes & Accessories
Reprocessing Do's and Don'ts
DO!
Wear utility gloves when handling contaminated cassettes and instruments.
Use Enzymax™ products when cleaning IMS™ Cassettes and instruments for best results.
4
4
Change the ultrasonic solution at least once a day or when visibly soiled. Dirty solution harbors particles that may inhibit efficient cleaning process. Wipe off cement residue, glass ionomer and/or other heavy debris at chairside immediately after use.
Sharpen instruments at chairside after sterilization and before a procedure. Sharpening Stones can be processed along with instruments in the IMS™ Cassette.
4
4
4
Autoclave utility gloves in seperate sterilization pouches.
4
DON’T!
Let instruments protrude from cassettes.
Reposition the location of the rails in cassettes. Improper rail location may cause undue stress on the cassette and can damage it.
6
6
Overload the cassette. If there is resistance when closing cassette, stop. Remove excess items and repeat the closing process. Forcing cassette to close under stress may inhibit future performance.
6
Use IMS™ Autoclave Wrap in dry heat sterilization.
6
Run ultrasonic unit or steam autoclave without water
6
Reprocess mixed metals in the same cycle.
6
B137