HuFriedyGroup | Product Catalog

Reference Guide

Crown Remover (CRL, CRU) Processing: • Clean, disinfect and sterilize in a completely disassembled state. Cleaning / Disinfection: • Do not disinfect with phenols or iodophors.

Osteotomes and Osteotome Handles Processing: • Clean, disinfect and sterilize in a completely disassembled state if applicable. Plastic Filling Instruments Processing: • Process in cassettes or trays with instrument rails to avoid scratches on the surface from other pointed instruments. Maintenance: • Residues of Filling Materials and etching products must be removed immediately. • Plastic Filling Instruments are designed with an extra smooth surface, in order to provide a better handling with composite materials. Scratches that are not visible might cause composite materials to stick to the rougher surface. Resin Instruments, Resin Components or Resin Cassettes Cleaning / Disinfection: • For resin or silicone products do not use detergents or disinfectants containing phenols or iodophors. Sterilization: • Dry Heat is explicitly not compatible with Instruments with resin handles (handle #8), with resin or Silicone components, inserts on any instruments, or with resin cassettes.

Root Canal Instruments Processing: • Reprocess in suitable endodontic stands (i.e. Hu-Friedy IMS-1275). Cleaning / Disinfection: • Pre-treatment should be conducted outside the Endodontic stand. • Automated cleaning and disinfection in a Washer-Disinfector unit is recommended. • Ultrasonic cleaning in the Endodontic stand is not recommended.

Sterilization: • Do not sterilize with dry heat.

Implacare Sterilization: • IMPLACARE disposable resin tips can be steam sterilized prior to use. • They are intended for one use only! • Clean, disinfect and sterilize in a completely disassembled state.

Scalpel Handles Processing: • Clean, disinfect and sterilize in a completely disassembled state.

Scaler marked with Color Coding Rings

Mallet Processing: • Clean, disinfect and sterilize in a completely disassembled state.

(IMS-1280L, IMS-1286L, IMS-1281, IMS-1287, IMS-12810, IMS-1287L, IMS-12810L, IMS- 1288, IMS-12811, IMS- 1288L, IMS-12811L, IMS-1289, IMS-1281L, IMS-1289L, IMS-1282, IMS-1282L, IMS-1283, IMS-1283L, IMS-1284, IMSö1284L, IMS-1234, IMS-1285, IMS- 1285L, IMS-1280, IMS-1286) Processing: • For reprocessing, attached Color Coding Rings do not have to be removed

Mouth Gags (MGA, MGC, MGI) Processing: • When using a cassette system for

cleaning/sterilization, the opening where the nylon tubing slips over the instrument tip must not be covered so as to allow the tips to properly drain. Clean, disinfect and sterilize in a completely disassembled state.

Syringes Processing: • Completely disassemble including unscrewing of the cylinder.

Retractors, Metal Processing: • Removable retractor tips must be disassembled from the handle before cleaning/disinfection and sterilization.

Mouth Mirrors Processing:

Ultrasonic Inserts, Magnetostrictive Processing: • Ultrasonic cleaning and disinfection as well as steam sterilization can be affected in suitable Hu-Friedy IMS-Cassettes. Cleaning / Disinfection: • For automated cleaning and disinfection in a Washer-Disinfector unit connecting rinsing adapters must be used, if the inserts are processed inside a cassette system. Otherwise open tray systems for the automated cleaning and disinfection or alternatively the manual cleaning and disinfection procedure are recommended.

• To avoid scratches on the mirror surface from other pointed instruments, reprocess in an instrument cassette with instrument rails. • Clean, disinfect and sterilize in a completely disassembled state. Cleaning / Disinfection: Note: All types of Rhodium coated Mouth Mirrors should not be cleaned and disinfected in an ultrasonic cleaner.

Retractors, Plastic (CRPC, CRPA) Cleaning / Disinfection: • Can only be disinfected by chemical disinfection. Do not clean / disinfect in Washer-Disinfector unit. Sterilization: • Do not sterilize (steam, dry heat etc.)!

O-Rings Sterilization: • O-Rings cannot be dry heat sterilized