EDVOTEK® Workshops - 2024 NSTA Denver

04 - Put the M into STEM: Quantitative Techniques for Biotechnology

Step 1: Agarose Gel Electrophoresis

RUNNING THE GEL

8. PLACE the gel (still on the tray*) into the electrophoresis chamber. COVER the gel with 1X Electrophoresis Buffer (See Table B for recommended volumes). The gel should be completely submerged.

9. PUNCTURE the foil overlay of the QuickStrip™ with a pipet tip. LOAD the entire sample (35 µL) into the well in the order indicated by Table 1, at right. 10. PLACE safety cover on the unit. CHECK that the gel is properly ori- ented. Remember, the DNA samples will migrate toward the posi- tive (red) electrode. 11. CONNECT leads to the power source and

REMINDER: Before loading the samples, make sure the gel is properly oriented in the apparatus chamber.

PERFORM electrophoresis (See Table C for time and voltage guidelines). Allow the tracking dye to migrate at least 3 cm from the wells. 12. After electrophoresis is complete, RE- MOVE the gel and casting tray from the electrophoresis chamber and PROCEED to gel staining.

Tube A Tube B Tube C Tube D Tube E Tube F TABLE 1: GEL LOADING 10 cycles 20 cycles 30 cycles

Lane 1 2 3 4 5 6

DNA Standard Marker 0 cycles 5 cycles

Time and Voltage Guidelines (0.8% Agarose Gel)

1x Electrophoresis Buffer (Chamber Buffer)

EDGE™

M12 & M36

Volts

Min/Max (minutes)

Min/Max (minutes)

EDGE™

150 mL

3 mL

147 mL

150 125 100

10/20

20/35 30/45 40/60

M12

400 mL

8 mL

392 mL

N/A

M36

1000 mL

20 mL

980 mL

15/25

45

1.800.EDVOTEK • www.edvotek.com • info@edvotek.com

Made with FlippingBook flipbook maker