STAT IM 2000 G4 Specifications p/n G4-121101 Unit size (L x W x H): 19.5” x 16.25” x 6” / 49.5 cm x 41.5 cm x 15 cm Cassette internal dimensions:
Rubber / Plastic
Unwrapped
Wrapped
Air Drying
3.5 minutes
10 minutes
15 minutes
Air drying only
Sterilization Time
11” x 7” x 1.4” / 28 cm x 18 cm x 3.5 cm Minimum shelf height for proper leveling: 8.5” / 22 cm Weight without water: 46 lbs / 21 kg Power consumption: 115 V, 60 Hz, 11A
135º C
135º C
121º C
Air drying only
Sterilization Temp
6 minutes
14 minutes
20 minutes
Air drying only
Total Cycle Time*
*Cycle time is dependent on instrument load and does not include drying.
STAT IM 5000 G4 Specifications p/n G4-201103 Unit size (L x W x H): 23.6” x 16.1” x 7.5” / 60 cm x 41 cm x 19 cm Cassette internal dimensions: 15” x 7” x 3” / 38 cm x 18 cm x 7.5 cm Minimum shelf height for proper leveling: 10” / 26 cm Weight without water: 73 lbs / 33 kg Power consumption: 115 V, 60 Hz, 11A
Rubber / Plastic
Unwrapped
Wrapped
Heavy Duty Unwrapped
3.5 minutes
6 minutes
35 minutes
6 minutes
Sterilization Time
132º C
132º C
121º C
132º C
Sterilization Temp
9 minutes
17.5 minutes
45 minutes
12 minutes
Total Cycle Time*
*Cycle time is dependent on instrument load and does not include drying.
VistaPure ™ Water Filtration
VistaPure ™ AutoFill Replenish any new or older-model STAT IM and BRAVO autoclave with distilled-quality water automatically, enabling truly hands-free convenience.
VistaCool ™ Direct-to-Drain System
Provides two grades of high-quality water for autoclaves, dental bottles, ultrasonic cleaners and the rinse cycle of instrument washers. V3000
Protects cabinets and drains—and saves valuable staff time—by cooling autoclave wastewater and sending it directly down the drain automatically.
S3020 - AutoFill for STATIM S3030 - AutoFill for BRAVO
V7501 - For One Autoclave V7502 - For Two Autoclaves
Manufactured by: SciCan Ltd. 1440 Don Mills Rd., Toronto, ON, Canada, M3B 3P9 Tel: 416-445-1600 Toll free: 1-800-667-7733 customerservice@scican.com Distributed by: SciCan Inc. 701 Technology Dr., Canonsburg, PA 15317, USA Tel: 724-820-1600 Toll Free: 1-800-572-1211 uscustomerservice@scican.com VistaPure ™ , VistaCool ™ and DrainHub ™ are manufactured by: Vista Research Group PO Box 321, Ashland, Ohio 44805-0321 USA Tel: (419) 285-3927 VistaResearchGroup.com
ENSURE ™ Monitoring for steam sterilization An easy to use, in-office monitoring system for steam sterilization. SCI-BISK-120 - Ensure BI Starter Kit (25 BI’s, Incubator & Log Book)
STATMAT ™ Protect your STAT IM
Securely seat cassettes atop your STAT IM with a safe and hygienic silicone mat .
01-112867S - For STATIM 2000 01-112868S - For STATIM 5000
www.scicanusa.com
Made with FlippingBook Digital Publishing Software