Maintaining the VistaPure system is simple and requires only periodic checks of water quality (with the provided TDS meters) and appropriate filter changes.

Please note: it is NOT typically necessary to change all of the VistaPure filters at the same time. The Prefilter and Polishing filter should be changed at least once per year, and the Deionization and Hyperfiltration filters changed only as needed, as indicated by TDS test.

Filter Position & Replacement Thresholds

System Flow

Prefilter (R3251) Replace this filter annually or as needed based on local water quality. Deionization Filter (R3252) Replace this filter and R3253 when system-mounted TDS meter “OUT” displays number ≥ 4 ppm. Deionization Filter (R3253) Replace this filter and R3252 when system-mounted TDS meter “OUT” displays number ≥ 4 ppm. Hyperfiltration Element (R3254) Replace whenever system-mounted TDS meter “IN” displays number > 10% of source water TDS. Polishing Filter (R3255) Replace this filter annually or as needed based on local water quality.

1

To autoclave wand (Deionized, distilled-quality water) 1 4 TANK 3

2

2

To faucet (Hyperfiltered, bottle-quality water) 1 4 5 TANK

3

FAUCET

4

To HYDR IM final rinse cycle (Hyperfiltered water)* 1 4 TANK

5

HYDR IM

* Requires optional Instrument Washer Adapter Kit (S3010)

Average Filter Replacement Cycle

Filter

12 mo

24 mo

36 mo

48 mo

60 mo

R3251

X

X

X

X

X

R3252*

X

X

X

R3253*

X

X

X

R3254

X

R3255

X

X

X

X

X

*  Assumes deionization filters will be replaced every 18 months on average. Ultimately, actual filter lifespan depends upon local water quality, inbound water pressure and volume of water used.

Manufactured by: Crosstex International, Inc. 1733 W. Parkside Lane Phoenix, AZ 85027 / USA T +1 (800) 531-2221 VistaResearchGroup.com

Distributed by: SciCan US Office 701 Technology Dr., Canonsburg, PA 15317 / USA TF +1 800-572-1211 T +1 724-820-1600 F +1 724-820-1479

www.scican.com

VistaPure and VistaCool are trademarks of Crosstex International, Inc., Phoenix, AZ., a Cantel Medical Company HYDRIM and STATIM are registered trademarks and BRAVO and STATCLAVE are trademarks of SciCan Ltd.