PUR® Surgical Protocol

PUR®

STANDARD PLACEMENT

IMPLANT DRIVER

The PUR® Implant Driver is used to drive all PUR® implants. Insert at 15-30 RPM, with torque not to exceed 55 Ncm. Once the implant is completely seated, pull to remove the driver from the implant.

PUR® implants are packaged without an Implant Mount attached. The PUR® Implant Drivers are used to carry the implant from the package to the surgical site and to drive the implant into the bone. Simply press the appropriate driver into the implant until the hex area on the driver engages the hex inside the implant.

Fully seated PUR® Implant

COVER SCREW

HEALING ABUTMENT

The appropriate PUR® Cover Screw may be used if this is to be a two-stage surgery. Tighten the cover screw with a .050” hex driver.

If this is to be a one-stage surgery, the appropriate PUR® Healing Abutment is placed and tightened using a .050” hex driver.