STANDARD PLACEMENT
TRU® DRILL SEQUENCE
Implant Diameter
3.5mm body diameter
4.3mm body diameter
5.0mm body diameter
6.0mm body diameter
TRU ® Drill Sequence TRU ® Drill Sequence Recommended Speed 1,000-1,500 RPM (Paralleling Pin is for guide only )
Bone Taps (3.5, 4.3, 5.0, 6.0mm)
Implant Driver Hex Drivers
Recommended Speed 15-30 RPM
Recommended torque not to exceed 55Ncm
TRU® DRILL DEPTH
FINAL DRILLING STEPS
Final Drilling step when placing the 3.5mm TRU® Implant Enlarge the osteotomy using the 3.0mm drill. If the bone is very dense, create internal threads in the osteotomy. Use the unique TRU® 3.5mm Titanium Bone Tap at 15-30 RPM. Final Drilling step when placing the 4.3mm TRU® Implant Enlarge the osteotomy using the 3.8mm drill. If the bone is very dense, create internal threads in the osteotomy. Use the unique TRU® 4.3mm Titanium Bone Tap at 15-30 RPM. Final Drilling step when placing the 5.0mm TRU® Implant Enlarge the osteotomy using the 4.5mm drill. If the bone is very dense, create internal threads in the osteotomy. Use the unique TRU® 5.0mm Titanium Bone Tap at 15-30 RPM. Final Drilling step when placing the 6.0mm TRU® Implant Enlarge the osteotomy using the 5.5mm drill. If the bone is very dense, create internal threads in the osteotomy using the unique TRU® 6.0mm Titanium Bone Tap at 15-30 RPM.
Note: The TRU® and PUR® drills are not true-to -depth, the drill tip should not be included in determining the final depth of the osteotomy in coordination with indicator lines on each drill.