TRU®
TRU® Surgical Protocol
STANDARD PLACEMENT
IMPLANT DRIVER
The TRU® Implant Driver is used to drive all TRU® implants. Insert at 15-30 RPM, with torque not to exceed 55 Ncm. Once the implant is completely seated, pull to remove the driver from the implant.
TRU® implants are packaged without an Implant Mount attached. The TRU® Implant Drivers are used to carry the implant from the package to the surgical site and to drive the implant into the bone. Simply press the appropriate driver into the implant until the hex area on the driver engages the hex inside the implant.
Fully seated TRU® Implant
COVER SCREW
HEALING ABUTMENT
The appropriate TRU® Cover Screw may be used if this is to be a two-stage surgery. Tighten the cover screw with a .050” hex driver
If this is to be a one-stage surgery, the appropriate TRU® Healing Abutment is placed and tightened using a .050” hex driver.
34
DENTAL IMPLANTS
WWW.STERNGOLD.COM