666273 Rev I Dental Implant Catalog

Guided Surgery

TRU® and PUR® Implant Digital Workflow

IMPLANT INSERTION 1. Using the guided driver of choice from the Sterngold Guided Surgery kit (Guided Latch, Guided Wrench, or Guided Mount Driver), install the TRU® & PUR® implants through surgical guide into the prepared osteotomy until the stop on the carrier/mount bottoms out. Recommended speed 15 – 30 RPM. It is important to stop rotating the implant once the stop contacts the guide sleeve to avoid stripping the bone and any loss in primary stability. Using the graduated torque wrench, confirm primary stability of 35 N-cm, not to exceed 55 N-cm. (Insertion Torque Value should be at 45 N-cm and Reverse Torque Value should be at 50 N-cm.)

2. Use the Bone Mill with Guide Pins to profile and prepare restorative platform area of implant site. (Recommended speed 200 RPM.)

3. Proceed to healing or restoration phase based on treatment plan.

COVER SCREW

HEALING ABUTMENT

54

DENTAL IMPLANTS

WWW.STERNGOLD.COM