IMPLANTS

PART # DESCRIPTION 901480 MOR ® Implant, 2.1 x 10mm 901481 MOR ® Implant, 2.1 x 13mm 901482 MOR ® Implant, 2.1 x 15mm 901488 MOR ® Implant, 2.1 x 18mm 901483 MOR ® Implant, 2.4 x 10mm 901484 MOR ® Implant, 2.4 x 13mm 901485 MOR ® Implant, 2.4 x 15mm 901489 MOR ® Implant, 2.4 x 18mm 901490 MOR ® Implant, 3.0 x 10mm 901491 MOR ® Implant, 3.0 x 13mm 901492 MOR ® Implant, 3.0 x 15mm 901493 MOR ® Implant, 3.0 x 18mm 904788 MOR ® RC-1 904814 MOR ® RC-1 (10-pack) 904789 MOR ® RC-2 904815 MOR ® RC-2 (10-pack) 904786 MOR ® Implant Analog 904787 MOR ® Impression Coping

MOR ® Mini Implant System Surgical Procedure

2.1 mm

3 Use the MOR ® Insertion Tool to deliver the implant from the package to the implant site and to begin hand driving the implant into the bone.

2 Peel back the cover on the inner blister package and upright the titanium tube that contains the MOR ® Implant.

1 The MOR ® 1.1mm Pilot Drill is designed to penetrate through the gingiva and cortical plate with an osteotomy about one-third to one-half the length of the implant for 2.1mm and 2.4mm implants. 2.0mm drill used for 3.0mm implant. A small flap or tissue punch is optional. Optional 2.7mm drill may be used in dense bone to widen osteotomy.

2.4 mm

3.0 mm

PROSTHETICS

4 When torque resistance is encoun- tered with the Insertion Tool, you may switch to the MOR ® Wing Driver .

5 As an option, you can use your hand- piece to complete the insertion of the MOR ® Implant using the MOR ® Latch Driver- Captive at 15-30 RPMs with irrigation and a torque setting no higher than 55 Ncm.

6 The final turn or two of the implant placement may be done by hand, using a torque wrench at a maximum value of 55 Ncm.

909014 O-ring Black for 1.8mm ball (10-pack)

909016 O-ring Purple Xtra Strong for 1.8mm ball (10-pack)

Operatory Placement of the MOR ® Retaining Caps for Overdenture Stabilization:

909013 Blockout Spacers (10-pack)

909017 Blockout Shims (25-pack)

904754 MOR ® Temporization Cap

Blockout shim

904819 MOR ® Comfort Cap (4-pack)

7 A block out spacer or blockout shim should be used to block out any exposed areas of the implant. Be sure to place right below the ball of the implant . Small rubber dam squares can also be used. Snap a Retaining Cap onto each implant.

8 Using a bur, prepare a recess in the denture over each Retaining Cap. The denture should not touch the Cap or it will not be seated properly on the tissue. Prime each recess with a thin coat of Stern Vantage Varnish LC, wait 30 seconds and light cure .

9 Carefully add composite (EZ PickUp ® ) over the top and sides of the cap. Place additional resin in the recesses in the overdenture and seat the prosthesis passively in the mouth. If the tissue is displaced it will make accurate seating of the caps very difficult.

904811 MOR ® Comfort Cap (10-pack)

904753 MOR ® Scan Body

905008 MOR ® Alignment Guide

904397 MOR ® Digital Spacer

INSTRUMENTS

905257 MOR ® Surgical Kit

905268 1.1mm MOR ® Drill 905288 2.0mm Latch Drill 905219 2.7mm Latch Drill

10 A fter resin has cured, remove denture. The MOR ® Retaining Caps will be picked up in the denture. O-rings used in MOR ® Retaining Caps are black. Clean and remove any excess resin.

905266 MOR ® Insertion tool

905269 MOR ® Wing Driver

OPTIONAL

905009 MOR ® Latch Driver-Captive

905259 MOR ® Implant Socket

905738 Graduated Torque Wrench

CONSUMABLE PRODUCTS

EZ PickUp ® Syringe, Tips and Varnish - self- curing attachment processing material

220237

For a new denture to be fabricated after implant placement, place the MOR ® Impression Copings onto each implant and take a closed tray impression. Send impression to your lab with MOR ® Analogs.

It is recommended that MOR ® Comfort Caps be placed over the balls of the MOR ® implants if the implants are not immediately loaded, for the healing period determined by the clinician. After they shall be removed by the clinician before restoration.

Use the MOR ® digital spacers to scan and design your overdenture.

220233 EZ PickUp ® tips

QuickLine ® Introductory Kit — long-term soft relining material (refills purchased separately)

220180

220181 QuickLine ® refill only

822000 MOR ® Implant Demonstration Model

666085 Rev. K