3.4 Operational Values

HVR100 systems are factory calibrated under the standard laboratory environmental conditions shown below:

Specification

Calibration Requirement

Ambient Temperature

20°C ± 0.5°C (68°F ± 1°F)

Humidity

40-60% RH

Temperature rate of change

0.5°C (1°F) per hour

HVR100 systems can be safely operated under the following environmental conditions:

Environmental Condition

Operational Requirement

Ambient Temperature

20°C ± 3°C (68°F ± 5°F)

Humidity

< 90% RH

Calibration Temperature

20 ± 0.5°C (68 ± 1°F)

Allowable Operating Temp.

18-22°C (64-72°F), non- condensing

If the system is to be operated under environmental conditions that are substantially different from those shown above, the system should be recalibrated under the expected conditions. Users should also consider the material characteristics of the parts under inspection, in particular coefficients of thermal expansion. Numerical compensation might be required when measuring parts under conditions different from those controlling the stated dimensional specifications for these parts.