HVR100 Manual

4.5 Setting Up the System PC

Note: This manual includes information specific to a Shuttle system PC. In some instances, the HVR system might include a system PC from another vendor.

To set up the system PC, refer to “System Controls” figure in Section 5 and perform the following steps: 1. Connect the USB cable from the top of the system PC to the I/O panel on the HVR metrology unit. 2. Connect an additional USB cable from the top of the system PC to the I/O panel on the HVR metrology unit to support the camera signal. 3. Connect the power supply cable to the underside of the system PC. 4. Connect the monitor cable from the monitor to the underside of the system PC. 5. Ensure both the wireless keyboard and wireless mouse are equipped with batteries. 6. Press the on/off switch on the back of the mouse to turn on the mouse. 7. Press the on/off switch on the top of the keyboard to turn on the keyboard. 8. Power on the system PC. Ensure that the system PC boots up properly. 9. Verify that the preinstalled M3 software is functioning properly. Refer to the M3 software documentation for additional information.

4.6 Setting Up the Metrology Unit

To set up the metrology unit, refer to “Metrology Unit I/O Control Panel” figure in Section 5 and perform the following steps: 1. Ensure both USB cables from the system PC have been connected to the I/O panel. See “Setting Up the System PC” earlier in this section. 2. Connect the power cable to the I/O panel. 3. Connect the remaining end of the power cable to an appropriate power outlet. 4. Connect the DB9 connector for the pendant to the I/O panel. 5. Power on the unit by pressing the Power On/Off switch to the “I” position.

4.7 Verifying the System is Operational

To verify the system is operational, perform the following steps:

1. Launch the M3 software from the system PC. 2. Place a sample part on the system stage.

3. Do a single measurement on an artifact whose dimensions are comparable to those that will be measured. Verify that the measured dimensions are within the HVR’s specifications.

4.8 On-site Functional Test, Calibration and Training

All Starrett vision metrology systems are calibrated at the factory prior to shipment; however, it is possible that components might have moved during shipment. A complete functional test and calibration are recommended following physical installation. Professional system installation is normally provided by Starrett or by an authorized Starrett dealer for all new systems sold in North America. Installation includes equipment setup, on-site calibration and on-site operator training. While professional installation i s a