System Maintenance KineMic inspection and metrology systems have been designed for years of superior service. Periodic maintenance as outlined in this section should be performed to maintain the system in peak operating condition.  Perform a daily inspection to ensure that the system is operating correctly and that proper safety guidelines are being followed.  Periodically verify basic optical performance.  Periodically perform cleaning.  For metrology systems, schedule regular factory-authorized calibration and maintenance service to preserve proper function and accuracy.  Replace batteries when needed for wireless keyboard and wireless mouse. Have alkaline AA batteries on hand for the mouse and AAA batteries for the keyboard.

9.

The wireless mouse takes one AA battery.

The keyboard takes two AAA batteries.

9.1 Daily Inspections

On a daily basis, inspect your system for general safety and basic functionality:  Verify that the work area is clean, dry and free of debris. Remove any debris or loose items from around the system and metrology stage.  Verify that the electrical power cords are plugged into a grounded power source and are unobstructed.  Verify that temperature and humidity are within recommended ranges.  Allow the system to warm up to normal operating temperature before performing critical parts measurements. 9.2 Weekly or Monthly Maintenance On a weekly or monthly basis (based on experience), do the following:  Inspect the system for cleanliness. If dirty, follow cleaning instructions as stated below under "Cleaning."  Verify that the stage control mechanisms move freely. If it binds, call for service. KineMic lead screws use with plastic nuts. Do not apply any grease or other lubricant.  Check the system for calibration against a certified chrome-on-glass standard if critical dimensions are to be measured.

- 40 -

Made with FlippingBook - professional solution for displaying marketing and sales documents online