Steps to validate calibration: 1. Place the calibrated verification standard in one the six positions on the glass stage as shown above. Secure the standard with hot melt glue or other suitable retaining method so that stage translation cannot move the standard under any condition. 2. Skew the center of the two end circles. Please refer to the M3 software manual for proper alignment procedures. 3. Measure the distance between the 1st and 2nd, 1st and 3rd, 1st and 4th, 1st and 5th, and the 1st and 6th fiducials. Repeat these measurements 10 times. 4. Calculate the absolute average deviation for each of the 5 groups of distance measurements. 5. Repeat the measurements for all six locations shown in the illustration. 6. The absolute averages should be within factory system specifications, and be derated for the environment and calibration errors.
Calibration Error Chart Example
9.5
Cleaning To the degree possible, KineMic systems should be kept in a clean environment, away from dirt, dust, oil and debris which could affect system performance or degrade the system’s mechanical and electronic parts. If a clean environment is not available, the machine should be kept as clean and protected as is possible. Cover machine with plastic when not in use. In harsh environments, preventive maintenance and factory service should be scheduled more frequently. 9.5.1 Cleaning External Surfaces Wipe down with a clean, lint-free cloth moistened (not wet) with plain water or Simple Green ® . Never wipe down with acetone or other harsh solvents, which may damage painted or plastic surfaces. Isopropyl alcohol may be used to clean surface contaminants where Simple Green proves ineffective.
- 43 -
Made with FlippingBook - professional solution for displaying marketing and sales documents online