8. Maintaining the System

HDV systems have been designed for years of superior service. Periodic maintenance outlined in this section should be performed to maintain the system in peak operating condition.

This section provides the following information:

Performing daily inspections

• Performing weekly or monthly inspections

Replacing the fan filter

Maintaining the inline clean air kit

• Performing monthly zoom optics verification

Performing calibration verification

Cleaning the system

Changing fuses

8.1 Performing Daily Inspections

On a daily basis inspect your system for general safety and basic functionality:

• Verify that the work area is clean, dry and free of debris. Remove any debris or loose items from around the system and stage. • Verify that the electrical power cord is plugged into a grounded power source and is unobstructed.

• Verify that temperature and humidity are within recommended ranges.

• Allow the system to warm up to normal operating temperature before performing critical parts measurements.

8.2 Performing Weekly or Monthly Maintenance

On a weekly or monthly basis (based on experience), do the following:

• Inspect the fan filter. If dirty, wash or replace the filter.

• Inspect the system for cleanliness. If dirty, follow cleaning instructions later in this section. • Verify that the stage control mechanisms move freely. Lubricate as needed or every six months as described later in this section.

• Check the system for calibration against a chrome-on-glass standard.