2. Environmental, Regulatory, and Safety Information
This section provides the environmental and safety information for the HDV systems.
2.1 Calibration and Operational Environment
HDV systems are factory calibrated under the standard laboratory environmental conditions detailed in the following table.
Specification
Requirement
Calibration Temperature
20 ° C ± 0.5°C (68°F ± 2°F)
Humidity
40-60% RH
Temperature rate of change
0.5°C (1°F) per hour
Allowable Operating Temperature
18-22°C (64-72°F), non-condensing
Environmental Protection
Positive pressure in enclosure provided by two fans with air filter
If the system is to be operated under environmental conditions that are substantially different from those shown above, the system should be recalibrated under the expected conditions. Users should also consider the material characteristics of the parts under inspection, especially particular coefficients of thermal expansion. Numerical compensation might be required when measuring parts under conditions different from those controlling the stated dimensional specifications for these parts.
Before performing critical parts measurements, allow the system to warm up and temperature stabilize for 30 minutes.
2.2 Regulatory Compliance
HDV300 and HDV400 metrology systems have been tested to comply with Council Directives 2006/42/EC. Systems that are operated in accordance with this guide carry the CE mark. The Product Safety, EMC Testing and Evaluations were provided by International Compliance Engineering (ICE), San Juan Capistrano, California, USA. The systems are also certified to be free from hazardous materials per comply 2002/95/EC RoHS.