2 Environmental and Regulatory Specifications
The following information provides the environmental and safety information for the HDV system.
2.1 Operating Environment Specifications
The following table provides the environmental specifications for the HDV system.
Feature
Specification
Calibration Temperature
20 ± 0.5°C (68 ± 1°F)
Allowable Operating Temp. 18-22°C (64-72°F), non-condensing
Positive pressure in enclosure provided by two fans with air filter
Environmental Protection
2.2 Calibration Environment Specifications
HDV500 systems are factory calibrated under the standard laboratory environmental conditions shown in the following table.
If the system is to be operated under environmental conditions that are substantially different from those provided, the system should be recalibrated under the expected conditions. Users should also consider the material characteristics of the parts under inspection, in particular coefficients of thermal expansion. Numerical compensation might be required when measuring parts under conditions different from those controlling the stated dimensional specifications for these parts.
Specification
Requirement
Ambient Temperature
20°C ± 0.5°C (68°F ± 1°F)
Humidity
40-60% RH
Temperature rate of change
0.5°C (1°F) per hour
2.3 Regulatory Compliance
The following information provides the regulatory and safety standards for the HDV500 system.
HDV500 metrology and inspection systems have been designed to comply with Council Directives 2006/42/EC - Machinery when installed and operated in accordance with this manual, and to carry the CE mark. HDV500 systems are certified by SKE to be free from hazardous materials per 2002/ 95/EC RoHS.